ST2523 Series Operation Manual Ver1.2
16
3.5
<BIN SETUP>
Press
BIN SETUP
in soft key area to enter into the
<BIN SETUP>
page shown in figure 3-6.
Figure 3-6 Limit Setup
Comparator functions can be set in this page. ST2523 can set 9 groups of nominal and limit values for primary
and secondary parameters.
Introduction to <BIN SETUP>
MODE
: Move the cursor to it, %TOL and ABS TOL will be displayed in the soft key zone.
When %TOL is selected, upper and lower limits will be displayed in the percentage error mode; When
ABS TOL is selected, upper and lower limits will be displayed in the absolute value error mode.
COMP
: COMPA and COMPB for primary and secondary parameters. COMPA (ON: the bin
comparison function of primary parameter will be enabled; OFF: the bin comparison function of
primary parameter will be disabled) and COMPB (ON: the bin comparison function of secondary
parameter will be enabled; OFF: the bin comparison function of secondary parameter will be
disabled).
NOM.A /NOM.B:
Nominal value is only valid in
%TOL mode. Nominal value for primary and
secondary parameters (NOMA and NOMB). You can input nominal values of primary and secondary
parameters through the numerical keyboard.
HIGH/LOW limit
: High and low limit for primary and secondary parameters. According to different
modes, the value of high and low limit can be respectively percentage value (-100~100) and absolute
value (-10000~10000).