ST2523 Series Operation Manual Ver1.2
22
ON:
Enable the password protection function. The passwords are used to start up the instrument and to protect
files.
MODIFY:
This key is used to modify password.
————————————————————————
NOTE:
The default password is 2523.
————————————————————————
4.1.4
Bus Mode
Set the communication interface.
Operation Procedures:
1)
Move the cursor to
Bus Mode
, the following items will be displayed.
RS232C
GPIB (option)
USBTMC
USBCDC
2)
Press
RS232C
, the RS232C interface is selected.
3)
Press
GPIB
, the GPIB interface is selected.
——————————————————————————————————
NOTE:
GPIB is an option. Only when it is installed, this mode can be available.
——————————————————————————————————
4)
Press
USBTMC
, the USBTMC interface is selected. The instrument communicates with PC through
the USB interface (USB DEVICE) on the rear panel.
5)
Press
USBCDC
, the USBCDC interface will be selected. The instrument communicates with PC
through USB (USB DEVICE) virtual serial interface.
4.1.5
Baud Rate
Baud rate is used to select the baud rate for RS232 interface. There are 6 selectable baud rates: 9600, 19200,
28800, 38400, 96000, and 115200.
Move the cursor to
Baud Rate
, the following items will be displayed in soft key area.
INCR+:
Touch this key to increase the baud rate.
DECR-:
Touch this key to decrease the baud rate.
4.1.6
Bus Addr
Bus address is used to control and display the current GPIB interface bus address.
Move the cursor to
Bus Addr
, the following items will be displayed in the soft key zone.
INCR++
Press this key, the instrument increases the number of bus address by 5.
INCR+
Press this key, the instrument increases the number of bus address by 1.
DECR-
Press this key, the instrument decreases the number of bus address by 1.
DECR--
Press this key, this instrument decreases the number of bus address by 5.
The bus address number ranges from 0 to 31.