ST2523 Series Operation Manual Ver1.2
14
3.4
<MEAS SETUP>
Press SETUP to enter into the <
MEAS SETUP
> page shown in figure 3-5.
Figure 3-5 Measurement Setup Page
The following items can be set in the <MEAS SETUP> page.
FUNC R-X:
Move the cursor to it, and then you can choose the desired parameters in the soft key
zone.
Range
(R)
:AUTO
: AUTO, HOLD, ↑, ↓will be displayed in the soft key zone by touching this key.
Press one of them, resistance range will be automatically selected or locked.
Range
(VDC)
:A
UTO
:
Move the cursor to it
, AUTO, HOLD, 6V/60V (ST2523), 30V/300V
(ST2523A) will be displayed in the soft key zone
. Press one of them, voltage range will be
automatically selected or locked. The voltage range can be switched between
6V/60V
and
30V/300V
.
Speed:MED
: Move the cursor to it, FAST, MED, and SLOW will be displayed in the soft key zone.
By touching them, the measurement speed can be toggled among them.
Trig:INT
: Move the cursor to it, INT, MAN, EXT and BUS will be displayed in the soft key zone.
The corresponding functions are respectively auto measurement mode, manual trigger measurement
mode, external trigger measurement mode and bus trigger measurement mode.
AVG:1
: Enter the desired number through the numerical keyboard ranging from 1 to 255, and then
press X1 to finish inputting.
Vm:OFF:
Move the cursor to it, ON and OFF will be displayed in the soft key zone. The
measurement voltage across DUT will be displayed in the <MEAS DISP> page when ON is selected;
when OFF is selected, it will be turned off.
Im:OFF:
Move the cursor to it, ON and OFF will be displayed in the soft key zone. The measurement
current through DUT will be displayed in the <MEAS DISP> page when ON is selected; when OFF is
selected, it will be turned off.
DEV A:OFF:
Move the cursor to it, OFF,
△
ABS and
△
% will be displayed in the soft key zone.