ST1952 Operation Manual
66
:RANGe Commands
:[UPPer] <n>
Command syntax:
:CAPacitance:RANGe[:UPPer] <n>
Set measurement range for capacitance
Command Parameter:
<n> = 1n,10n,100n,1u,10u,100u,1000u,10000u)
Expected reading is F
DEFault
1nF
MINimum
1nF
MAXimum
10000µF
Query:
:RANGe[:UPPer]?
Query measurement range of the current function.
Description:
This command is used to manually select the measurement range for the specified measurement
function.
:AUTO <b>
Command syntax:
:CAPacitance:RANGe:AUTO <b>
Set auto range for capacitance
Command parameter:
<b> = 1 or ON
Enable auto range
0 or OFF
Disable auto range
Query:
:AUTO?
Query auto range (ON or OFF)
Description:
These commands are used to control auto ranging. With auto ranging enabled, the instrument
automatically goes to the most sensitive range to perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that manually selects the
measurement range (:RANGe <n>). When auto range is enabled, the parameter value for:RANGe
<n> changes to the automatically selected range value. Thus, when auto range is disabled, the
instrument remains at the automatically selected range. When a valid:RANGe <n> command is sent,
auto ranging disables.
:REFerence <n> Commands
:REFerence <n>
Command syntax: