Section G
AirWave System Operator’s Manual
40
c. Ready to Set Baseline
A flashing
Set Baseline Icon
is displayed indicating that the system is ready to set the baseline ETT
position and passageway size (see Figure G-47). Once the baseline is set, all movement of the ETT tip
will be measured from that point.
Figure G-47
Press the
Set Baseline Button
(
, middle function button as shown in Figure G-47). The system
asks to verify setting the baseline (see Figure G-48) and, upon verification, plays a confirmatory tone.
Figure G-48
d. Baseline Set
The baseline position and passageway size are set (see Figure G-49). All indicators are now fully
functional.
Figure G-49