Section G
AirWave System Operator’s Manual
25
g. Patient Info
Selecting
Patient Info
enters the
Patient Info Screen
(see Figure G-16) where the patient identification (ID)
number can be entered or changed (if the System Administrator has enabled this feature).
Figure G-16
1. Change the Patient ID as previously described in
Section F.6 - Basic Navigation and Editing
. Note
that each individual digit is selected using the left/right arrows and modified using the up/down arrows.
2. Once changes are complete, press the
Previous Screen
button to exit the
Patient Info Screen
. The
system prompts to accept or cancel changes.
3. Note that if Data Logging (see Section G.3.h) is active, changing the Patient ID will erase all
logged data.
h. Data Logging
Selecting
Data Logging
enters the
Data Logging Screen
, which provides options for collecting data when
troubleshooting or doing research. Detailed instructions can be found in the
Troubleshooting
Section of the
AirWave System Set-up and Maintenance Manual.
Please contact your Biomedical or Clinical Engineering Team
or SonarMed technical support for assistance.
i. System Settings
Selecting
System Settings
enters a password screen which leads to a set of screens where the LCD
brightness, Date-Time, Patient ID on/off, Alarm Sound selection, and Alarm default settings can be modified.
These settings are intended to be modified by authorized personnel only (See
AirWave System Set-up and
Maintenance Manual
).
4. Connecting an AirWave Sensor to an ETT
NOTICE! When possible, the AirWave Sensor should be connected to the ETT before patient intubation
as the System can provide useful information during the intubation process.
a. Select the correct Sensor size to match the ETT inner diameter. The size is found on the Sensor
package label and also on the Sensor serial number label (see
b. Figure G-17). See
AirWave Sensor IFU
for details.