Section L
AirWave System Operator’s Manual
53
Weight
5.5
– 9.0 mm Sensor:
1.0 oz / 28 g
2.5
– 5.0 mm Sensor:
0.5 oz / 14 g
Cable:
3.5 oz / 99 g
3. System Accuracy
Indicator
Accuracy
(95% Confidence)
ETT Tip Movement
± 1 mm
ETT Obstruction (%)
± 15%
1
Passageway Diameter (mm)
± 25%
2
1
– Percentage is absolute. For example, if actual obstruction is 50% of ETT cross-sectional area, then accuracy is 35% to 65%.
2
– Percentage of passageway diameter. For example, if actual diameter is 10.0 mm, then accuracy is 7.5 mm to 12.5 mm.
4. Storage & Operating Conditions
When not in use, the SonarMed AirWave is to be stored under the following conditions:
Sensor:
•
Temperature:
-20 to +50 °C
•
Relative Humidity:
10% to 95%
•
Atmospheric Pressure:
50 to 106 kPa
Monitor:
•
Temperature:
-40 to +70 °C
•
Relative Humidity:
10% to 95%
•
Atmospheric Pressure:
50 to 106 kPa
The Monitor and Sensors may be used under the following conditions:
•
Temperature:
+10 to +40 °C
•
Relative Humidity:
30% to 95%
•
Atmospheric Pressure:
70 to 106 kPa