Model 464 Electronic Pump Control Unit User Guide

Page 2

User Flow Rates
LCD Contrast
About

ECU 464-125 Main

Preset Flow Rates

4:59 Min

99%

 Timer

: indicates the amount of time remaining before the Control 

Unit automatically turns off. The Control Unit will automatically turn off after  

5 minutes of being idle.

User Flow Rates
LCD Contrast
About

ECU 464-125 Main

Preset Flow Rates

4:59 Min

99%

 Battery icon

: represents the battery life remaining. Battery life is also 

numerically displayed as a percentage to the left of the icon. The icon will flash 

when the battery percentage is 0.

99

User Flow Rates

LCD Contrast

About

ECU 464 Main Menu

Preset Flow Rates

 

99

User Flow Rates
LCD Contrast
About

ECU 464 Main Menu

Preset Flow Rates

 Arrows

: left pointing arrow indicates there is at least one menu that can 

be accessed using the left cursor key.  Right pointing arrow indicates a sub-menu 

exists for the active menu item, and can be accessed using the right cursor key 

or OK button (moves to next item).

Figure 1-2 Electronic Control Unit LCD Display

User Flow Rates
LCD Contrast
About

ECU 464-125 Main

Preset Flow Rates

4:59 Min

99%

OK

 OK button

: selects a highlighted menu item (also toggles between lower 

and uppercase letters and pressure units).  Press and hold for at least 3 seconds 

to turn the Control Unit off.  (However, this will not work in the Contrast Menu, 

you can only use the left cursor key to exit the menu, and the plus/minus keys 

to adjust the contrast.) 

 

 

 

 Cursor keys

: navigate through the menus and menu items.

 

 

 Plus/Minus keys

: cycle through numbers and letters when editing or 

creating new flow rates.  Increase/decrease LCD display contrast. (Hold down 

to scroll quickly through values).

Manual Control Valve

:

 

allows manual operation of the Control Unit.  When 

pushed the solenoid opens.  See Page 8.

Air In: 

connection for the supply line from the compressed gas supply source.  

(In-line filter is not required.)

Air Out

: connection for the drive line from the pump.  

Regulator

: sets the pumping pressure.  Turn clockwise to increase the pumping 

pressure and counter-clockwise to decrease pressure.  

Pressure Gauge: 

displays the pumping pressure.

Battery Enclosure

: houses four (4) AA alkaline batteries.  

1.4  LCD Display

1.3  Control Panel

  NOTE

Pressing any key will turn the 
Control Unit on (except Manual 
Control button).

  NOTE

The Control Unit will turn off 
automatically after 5 minutes of 
being idle, provided that cycling is 
not active. 

 

  NOTE

Battery Life is 40,000 drive/vent 
cycles  or 100 hours at 10 second 
drive/vent cycles.

Summary of Contents for 464 Mk3

Page 1: ...Model 464 Mk3 Electronic Pump Control Unit User Guide September 24 2019 109652 Rev C...

Page 2: ......

Page 3: ...Rates 5 2 5 1 Saved Settings 5 Edit Settings 6 Delete Settings 6 2 5 2 Create Setting 7 2 6 Automatic Drive Vent Cycles 7 2 7 Manual Drive Vent Cycles 8 2 8 Battery Replacement 8 3 0 Pumping Instruct...

Page 4: ......

Page 5: ...4 9 23 x 26 x 12 cm Drive Vent Time Range 1 999 seconds Maximum Site Name Length 16 characters upper or lowercase and numeric Memory Capacity 99 user flow rates in non volatile FRAM 3 preset in Flash...

Page 6: ...ghlighted menu item also toggles between lower and uppercase letters and pressure units Press and hold for at least 3 seconds to turn the Control Unit off However this will not work in the Contrast Me...

Page 7: ...1 905 873 2255 2 1 1 Main Menu There are four main menu items for the Electronic Control Unit Figure 2 2 Preset Flow Rates allows you to select a flow rate with predefined drive and vent cycles Low F...

Page 8: ...V2 000 Solinst Canada Ltd 1 905 873 2255 www solinst com About ECU 464 125 4 59 Min 99 Set LCD Contrast 50 4 59 Min 99 2 4 Preset Flow Rates Preset Flow Rates have predefined drive and vent times they...

Page 9: ...tings Selecting Saved Settings will display a list of all previously created user defined flow rates Figure 2 9 User Defined Saved Settings Use the cursor keys to navigate through the settings To rear...

Page 10: ...on the Unit Pressing OK toggles between psi and KPa The pressure is automatically converted to the new units The displayed pressure does not control the actual pressure it is only a reminder of the pr...

Page 11: ...99 2 6 Automatic Drive Vent Cycles Select the desired setting To start the drive vent cycles use the cursor key to highlight Start and press OK When the Control Unit is running the LCD will display th...

Page 12: ...n Battery Enclosure Pressure Gauge If the battery level is low the battery warning will be displayed after attempting to start a drive vent cycle Figure 2 16 The battery icon will also be flashing Pre...

Page 13: ...reel The Drive Line supplied with the Quick Exhaust fitting should always be used 4 Press any button on the keypad to turn the Control Unit on 5 Select the desired pumping settings 6 Set the compress...

Page 14: ...utton on the Electronic Control Unit Count the seconds until the bubbles stop forming in the water This count is your Drive Time as when the air bubbles stop the bladder is fully compressed Bladder Pu...

Page 15: ...Drive Time is 40 of this count i e the full count is the amount of time is takes to evacuate the entire pump system of water using drive gas 40 of that count means the drive gas will not enter the pum...

Page 16: ...increase the vent time then re adjust the drive time to obtain the highest flow rate Double Valve Pumps Select and start your desired flow rate from the User Flow Rates menu If a higher flow rate is r...

Page 17: ...to use the External Filter The External Filter connects to the 464 Supply Line and reduces the risk of moisture entering the Control Unit See separate operating and installation instructions for the 1...

Page 18: ...r use the draining valve as necessary to prevent water from entering the 464 Electronic Pump Control Unit Of note sources of moisture may include 1 The air compressor follow the manufacture s suggesti...

Page 19: ...ged with compressed air after use as described in Section 5 0 then stored in a dry condition Ensure the 464 External Filter is dry before storage If the Pump Control Unit is going to be stored for lon...

Page 20: ...Ltd 35 Todd Road Georgetown ON L7G 4R8 Fax 1 905 873 1992 800 516 9081 Tel 1 905 873 2255 800 661 2023 instruments solinst com www solinst com High Quality Groundwater and Surface Water Monitoring In...

Reviews: