Model 464 Electronic Pump Control Unit User Guide
Page 1
1.0 Introduction
1.1 Operating Principles
The Model 464 Electronic Control Unit controls the supply of compressed gas
to pneumatic pumps. Drive (pressure) and vent periods are cycled to provide
water flow. During pressurization periods, water is forced into the sample tubing.
The vent period allows water to re-enter (recharge) the pump. Cycle repetition
may be controlled manually or automatically using pre-set pumping rates.
1.2 Model 464 Electronic Control Unit Specifications
Figure 1-1 Model 464 Electronic Control Unit (125 psi)
Model 464 Specifications
125 psi
250 psi
Operating Temperature:
-20ºC to 50ºC
Weight:
5 lbs (2.3 Kg)
7.8 lbs (3.5 Kg)
Dimensions:
9.1 x 10.2 x 4.9” (23 x 26 x 12 cm)
Drive/Vent Time Range:
1-999 seconds
Maximum Site Name
Length:
16 characters (upper or lowercase and numeric)
Memory Capacity:
99 user flow rates in non-volatile FRAM
3 preset in Flash
Battery Life:
40,000 drive/vent cycles @ 25ºC from 4 AA alkaline batteries
(100 hours based on 10 second drive/vent cycles)
Drive Gas Sources:
Nitrogen Cylinder, Compressed Air Cylinder, Air Compressor
Supply/Drive Line Connections:
M Style Quick-Connect 1/4" NPT
Maximum Output Pressure:
125 psi (861 KPa)
250 psi (1724 KPa)
Maximum Input Pressure:
150 psi (1034 KPa)
300 psi (2068 KPa)
Manual Control
Button
Battery
Enclosure
Regulator
Pressure Gauge
Air Out
Air In
Drive Line
Supply Line
NOTE
A Nitrogen cylinder is ideal as it is
an inert, dry gas, best for pristine
samples and longer life expectancy
of the Control Unit.
Pneumatic Pump
Operating Principles:
Scan to View
407 Bladder Pump Video >
Scan to View
408 Double Valve Pump Video >
Summary of Contents for 464 Mk3