background image

Model 464 Electronic Pump Control Unit User Guide

Page 6

Delete Settings

To delete the setting, use the left arrow key to highlight the delete icon   and 

press OK.  After pressing OK, a prompt will be displayed to confirm the deletion.

Figure 2-10 Delete Saved Settings

W67 Farm 12
E89 Spill 2
A22 VOC 9

Saved Settings

W13 Landfill 34

Delete site
“W13 Landfill 34”?

Yes

No

4:59 Min

99%

Edit Settings

Select a particular setting by pressing the right arrow or OK.

Figure 2-11 Saved Setting Information

To edit site information, use the cursor keys to navigate the fields and modify 

them using the +/- keys.  Holding the +/- keys will cycle through letters and 

numbers quickly.  The OK button toggles between upper and lower case letters.  

The “Site” field has a fixed length of 16 characters.  

Use the +/- keys to change the drive and vent times.  The maximum drive/vent 

time is 999 seconds (16.65 minutes).

Use the +/- keys to change the pressure setting.  The maximum pressure is  

125 psi (861 KPa) or 250 psi (1724 KPa), depending on the Unit.  Pressing OK 

toggles between psi and KPa.  The pressure is automatically converted to the 

new units.  The displayed pressure does not control the actual pressure, it is only 

a reminder of the pressure to adjust externally.

When finished editing the site settings, use the cursor key to select “Save 

Changes” or “Start” to start pumping without saving the changes.  When “Stop” 

is selected, a display will appear asking if you would like to save changes.  User 

settings are saved to non-volatile memory so that they will be retained if the 

batteries are removed.  A maximum of 99 user settings are supported.

Figure 2-12 Editing Saved Setting Information

Site:    W13 Landfill 37
Drive: 

 5s   Set Pressure:

Vent:  5s   25psi

Saved Settings [1 of 10]

Start

4:59 Min

99%

Site:    W13 Landfill 3

7

Drive:   5s      Set Pressure:
Vent:   5s      25psi

Saved Settings [1 of 10]

Start

Save Changes

4:59 Min

99%

 

  NOTE

The maximum  drive/vent time 
is 999 seconds each and the 
maximum pressure setting is 125 
psi (861 KPa) or 250 psi (1724 KPa), 
depending on the Unit.

 

  NOTE

The displayed pressure does not 
control the actual pressure, it is only 
a reminder of the required output 
pressure for that site.

 

  NOTE

•  1 psi = 2.3 ft of water
•  1 KPa = 0.1 m of water

Summary of Contents for 464 Mk3

Page 1: ...Model 464 Mk3 Electronic Pump Control Unit User Guide September 24 2019 109652 Rev C...

Page 2: ......

Page 3: ...Rates 5 2 5 1 Saved Settings 5 Edit Settings 6 Delete Settings 6 2 5 2 Create Setting 7 2 6 Automatic Drive Vent Cycles 7 2 7 Manual Drive Vent Cycles 8 2 8 Battery Replacement 8 3 0 Pumping Instruct...

Page 4: ......

Page 5: ...4 9 23 x 26 x 12 cm Drive Vent Time Range 1 999 seconds Maximum Site Name Length 16 characters upper or lowercase and numeric Memory Capacity 99 user flow rates in non volatile FRAM 3 preset in Flash...

Page 6: ...ghlighted menu item also toggles between lower and uppercase letters and pressure units Press and hold for at least 3 seconds to turn the Control Unit off However this will not work in the Contrast Me...

Page 7: ...1 905 873 2255 2 1 1 Main Menu There are four main menu items for the Electronic Control Unit Figure 2 2 Preset Flow Rates allows you to select a flow rate with predefined drive and vent cycles Low F...

Page 8: ...V2 000 Solinst Canada Ltd 1 905 873 2255 www solinst com About ECU 464 125 4 59 Min 99 Set LCD Contrast 50 4 59 Min 99 2 4 Preset Flow Rates Preset Flow Rates have predefined drive and vent times they...

Page 9: ...tings Selecting Saved Settings will display a list of all previously created user defined flow rates Figure 2 9 User Defined Saved Settings Use the cursor keys to navigate through the settings To rear...

Page 10: ...on the Unit Pressing OK toggles between psi and KPa The pressure is automatically converted to the new units The displayed pressure does not control the actual pressure it is only a reminder of the pr...

Page 11: ...99 2 6 Automatic Drive Vent Cycles Select the desired setting To start the drive vent cycles use the cursor key to highlight Start and press OK When the Control Unit is running the LCD will display th...

Page 12: ...n Battery Enclosure Pressure Gauge If the battery level is low the battery warning will be displayed after attempting to start a drive vent cycle Figure 2 16 The battery icon will also be flashing Pre...

Page 13: ...reel The Drive Line supplied with the Quick Exhaust fitting should always be used 4 Press any button on the keypad to turn the Control Unit on 5 Select the desired pumping settings 6 Set the compress...

Page 14: ...utton on the Electronic Control Unit Count the seconds until the bubbles stop forming in the water This count is your Drive Time as when the air bubbles stop the bladder is fully compressed Bladder Pu...

Page 15: ...Drive Time is 40 of this count i e the full count is the amount of time is takes to evacuate the entire pump system of water using drive gas 40 of that count means the drive gas will not enter the pum...

Page 16: ...increase the vent time then re adjust the drive time to obtain the highest flow rate Double Valve Pumps Select and start your desired flow rate from the User Flow Rates menu If a higher flow rate is r...

Page 17: ...to use the External Filter The External Filter connects to the 464 Supply Line and reduces the risk of moisture entering the Control Unit See separate operating and installation instructions for the 1...

Page 18: ...r use the draining valve as necessary to prevent water from entering the 464 Electronic Pump Control Unit Of note sources of moisture may include 1 The air compressor follow the manufacture s suggesti...

Page 19: ...ged with compressed air after use as described in Section 5 0 then stored in a dry condition Ensure the 464 External Filter is dry before storage If the Pump Control Unit is going to be stored for lon...

Page 20: ...Ltd 35 Todd Road Georgetown ON L7G 4R8 Fax 1 905 873 1992 800 516 9081 Tel 1 905 873 2255 800 661 2023 instruments solinst com www solinst com High Quality Groundwater and Surface Water Monitoring In...

Reviews: