CADD
®
-Solis VIP Pump
p. 27
3.
Close the battery door and using
your fingers, the pump key, or a
coin, turn the knob on the battery
door clockwise to lock it
Note: If you put the batteries in
backwards, the pump will not power
up. Check the batteries, making sure to
match the + and – markings.
AA batteries
with AC adapter
Rechargeable
battery pack
Rechargeable
battery with
AC adapter
AA batteries
REPLACING THE BATTERY DOOR
If the battery door
is removed or
needs replacing,
simply snap the
door onto the bar
that is located on
the pump.
POWER UP
Press and hold the power switch to
turn the pump on. The pump starts
the power up sequence during which it
performs various self-tests and tests
for alarm conditions.
•
If any issues are found while the
pump is performing the self tests,
alarms will sound
•
If you believe there is a problem,
remove the pump from service
and contact Smiths Medical
Customer Service
CAUTION: If the power up results
in an error message indicating that
the protocol library was lost, do
not proceed with using the pump.
Follow your facility’s procedures for
downloading protocol libraries.
POWER DOWN
Press the power switch to turn the
pump off. You must acknowledge the
“Power down?” prompt by selecting
Yes. The pump then powers down.
CASSETTES
The cassette is part of the CADD™
medication cassette reservoir or
CADD
®
administration set that attaches
to the bottom of the pump.
Notes:
•
A CADD
®
administration or
extension set with free-flow
protection must be used
•
A CADD
®
high volume
administration set is required
for rates above 250 mL/hr