CADD
®
-Solis VIP Pump
p. 41
Alarm / Message
Alarm
Priority
Description / Corrective Action
Battery low.
Replace battery.
Low
The rechargeable battery pack or the 4 AA batteries are low
but the pump is still operable. Select Acknowledge to clear the
alarm, or the alarm will automatically clear after 5 seconds.
Recharge or change the rechargeable battery pack or replace
the 4 AA batteries soon.
Battery removed.
Pump will not run.
Medium
The pump is stopped and the rechargeable battery pack or the
4 AA batteries were removed, but the pump is still powered by
the AC adapter. Select Acknowledge to clear the alarm.
Install a fully charged rechargeable battery pack or 4 new AA
batteries. To start delivery, good batteries must always be
installed, even when an external source of power is connected.
Cannot start pump with
a depleted battery.
Medium
The battery power is too low to operate the pump. To start
delivery, good batteries must always be installed even when an
external source of power is connected. Select Acknowledge to
clear the alarm. Install 4 new AA batteries or a fully charged
rechargeable battery pack. If appropriate, start the pump.
Cannot start pump
with a reservoir
volume of zero.
Medium
The reservoir volume in the pump is set to zero. Select
Acknowledge to clear the alarm. Edit or reset the reservoir
volume to the correct value. If appropriate, start the pump.
See page 32 for more information about resetting the
reservoir volume.
Cannot start
pump without a
latched cassette.
Medium
The pump will not start without a cassette attached. Select
Acknowledge to clear the alarm. Make sure a cassette is
properly attached, then start the pump.
NOTE: This alarm has associated help screens.
Cassette not
attached properly.
Reattach cassette.
High
The cassette is not properly attached. Close the tubing,
remove the cassette, then reattach. If the alarm persists,
replace the cassette.
NOTE: You must remove the cassette to continue.
Cassette was partially
unlatched. Fully
remove and reattach
the cassette.
Medium
The cassette was not completely removed from the pump
before it was reattached and the pump’s sensors are not able
to detect the cassette type. Remove the cassette and reattach
it, then verify the cassette type in the pump display. If the alarm
persists, replace the cassette.
NOTE: You must remove the cassette to continue.
Check for empty
tubing or reservoir.
Pump stopped.
High
The tubing beneath the pump may not contain fluid, or the fluid
container may be empty. The pump is stopped and will not run.
Select Acknowledge to clear the alarm. Check to see if the fluid
container is empty. If fluid is present in the reservoir, clamp the
tubing, remove the cassette, and check for air in the tubing.
If the alarm persists, remove the pump from service and
contact Customer Service to return the pump for service.
Dose is now overdue.
Pump is stopped.
Medium
The pump is stopped and a dose is overdue for its scheduled
delivery. Select Acknowledge to clear the alarm, and then
start the pump.