Clinician Guide
p. 6
KEYPAD
2
1
6
3
4
2
5
1.
stop/start - Start and stops
pump delivery
2.
soft keys - Answer questions on the
pump’s display or navigate through
pump screens
3.
▼
- Navigates through the
pump menus, scrolls down
or decreases values
4.
▲
- Navigates through the
pump menus, scrolls up
or increases values
5.
select - Selects a menu item
6.
PCA dose - Requests a PCA
dose if the remote dose cord is
not connected and the PCA dose
option is available
USB Port - A mini-B USB cord can
be attached to the USB port for
communication with the CADD
®
-Solis
medication safety software.
Remote Dose Cord Jack - Used for
attaching the remote dose cord to
the pump.
Battery Compartment - Four AA
batteries or the rechargeable battery
pack fit into this compartment. The
batteries are the primary source of
power or serve as a back-up when the
AC adapter is in use.
Cassette Latch - Used to attach the
cassette to the pump.
Cassette / Keypad Lock - Secures the
cassette to the pump using the pump
key. The cassette latch must be closed
before it can be locked.
Air Detector - The air detector can be
turned on or off, depending on facility
or therapy requirements. If it is turned
on and air is detected in the part of
the tubing that passes the air detector
sensor, an alarm sounds and
delivery stops.
Downstream Occlusion Sensor - When
a downstream occlusion (between the
pump and the patient) is detected, an
alarm sounds, delivery stops and the
display shows a message.
Upstream Occlusion Sensor - When an
upstream occlusion (between the pump
and the reservoir) is detected, an alarm
sounds, delivery stops and the display
shows a message.