Smiths Medical CADD Solis v 3.0 Manual Download Page 8

Screen is blank and alarm is sounding.

Alarm Priority High. The pump has lost power and is 

no longer delivering. The pump was delivering and the 

batteries were removed or the battery door was opened. 

Clear this alarm by replacing the batteries or closing the 

battery door, then turn the pump back on. Or the alarm 

will stop after the power has been off for a minimum of 2 

minutes.

Air-in-line detected. Press “acknowledge” then 

prime tubing.

Alarm Priority High. The pump is stopped and can not 

run.The air detector has detected air in the fluid path; 

the fluid path may contain air bubbles. Acknowledge the 

alarm. Then, if the fluid path contains air bubbles, close 

the clamps, disconnect the fluid path from the patient, 

and follow the instructions for priming to remove the air.

Battery depleted. Pump stopped.

Alarm Priority High. Install 4 new AA batteries or a 

fully charged rechargeable battery pack. In order to 

start delivery, good batteries must always be installed, 

even when an external source of power is connected. If 

appropriate, restart the pump.

Battery low. Replace battery.

Alarm Priority Low. Charge (or change) the rechargeable 

battery pack or change the 4 AA batteries soon.

Current settings require high/standard volume set. 

Change cassette.

Alarm Priority High. A high volume or standard volume 

administration set is required. The pump is stopped and 

will not run. Remove the administration set to continue.

Delivery limit reached. Or, delivery limit reached  

and partial dose delivered.  

Troubleshooting

Pump’s status bar reads “KVO = 0”

Alarm Priority Low. The programmed delivery  

limit has been reached, and the pump is not delivering 

fluid. This alarm occurs when the continuous rate or a 

PCA dose has caused the delivery limit to be exceeded. 

Acknowledge the alarm (the alarm automatically 

clears after 5 seconds).

Pump’s status bar reads “Del Limit”

Alarm Priority Low. The programmed delivery limit 

has been reached, and the pump is delivering fluid at 

the KVO rate of 0.1mL/hr. This alarm occurs when the 

continuous rate or a PCA dose has caused the delivery 

limit to be exceeded. Acknowledge the alarm (the 

alarm automatically clears after 5 seconds).

Summary of Contents for CADD Solis v 3.0

Page 1: ...ator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J Cassette Keypad Lock K Power Switch CADD Solis v 3 0 Ambulatory Infusion System with Programmed Intermittent Bolus Quick Reference Card for Clinicians ...

Page 2: ...en press Save Note If the desired value is outside the soft limit confirm the soft limit override by pressing Yes Note The next bolus setting allows for a one time override of the intermittent bolus cycle as defined by the bolus interval 13 Continue until all patient specific parameters have been reviewed and or edited Press Accept Value for each setting A check mark appears next to each patient s...

Page 3: ...t soft key twice Tasks then Lock Keypad If a key was used turn the key clockwise to relock the cassette and keypad 4 Verify that the keypad and cassette are locked Changing a patient s current program while the pump is running Pump must be running 1 From the home screen Press Tasks 2 Give Clinician Bolus is highlighted Press Select 3 Enter the clinician security code 4 The screen displays the clin...

Page 4: ... Review pump settings displays Press Review 8 Choose Accept Value to confirm the value is correct for the highlighted patient specific parameter or press Select to edit the highlighted parameter 9 Continue until all patient specific parameters have been reviewed accepted and display checkmarks Press Next 10 Start Pump displays Press Yes Note If a security code was used to unlock the keypad the key...

Page 5: ...iew pump settings displays Press Review 10 Choose Accept Value to confirm the value is correct for the highlighted patient specific parameter or press Select to edit the highlighted parameter 11 Continue until all patient specific parameters have been reviewed accepted and display checkmarks Press Next 12 Start Pump displays Press Yes Note If a security code was used to unlock the keypad the pump ...

Page 6: ... Scroll or to the desired report and press Select 2 Press Back to return to the Reports menu and then press Back again to return to the home screen Viewing reports Option 2 1 From the home screen press Tasks Scroll to View Reports and press Select Scroll or to the desired report and press Select 2 Press Back to return to the Reports menu Press Back again to return to the home screen To clear Given...

Page 7: ...y an amber screen it continues until acknowledged or until the condition that triggered the alarm goes away Low Priority Alarm A low priority alarm does not stop the pump Accompanied by a blue screen the alarm automatically clears after 5 seconds or when the condition that triggered the alarm goes away Informational Message This alarm does not stop the pump This message appears in the status bar I...

Page 8: ...en an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Charge or change the rechargeable battery pack or change the 4 AA batteries soon Current settings require high standard volume set Change cassette Alarm Priority High A high volume or standard volume administration set is required The pump is stopped and will not run Remove th...

Page 9: ...Reservoir volume is zero Pump stopped Alarm Priority High The reservoir volume has reached 0 0 ml The pump stops and can not run Acknowledge the alarm Install a new fluid container Reset or edit the value of the reservoir volume Upstream occlusion Clear occlusion between pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be resulting from a k...

Page 10: ...y business Smiths Group plc Please see the Instructions for Use Operator s Manual for a complete listing of the indications contraindications warnings and precautions CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2017 Smiths Medical All rights reserved Prod...

Reviews: