Changing the IV bag without changing the tubing.
Stop the pump
1.
Press Stop/Start.
2.
“Stop Pump?” displays. Press Yes. Aseptically
remove the empty IV bag from the tubing and attach
the new IV bag.
Reset reservoir volume
3.
Scroll
until Reservoir Vol is highlighted.
Press Select.
4.
Screen displays “Reservoir Volume remaining: XXmL
Reset?” Press Yes.
5.
Unlock the keypad using the security code or the
pump key.
6.
The screen displays the current reservoir volume and
a scroll range.
7.
Press Select to reset the reservoir volume to the
amount displayed in blue text or scroll
or
to
adjust the value. Press Save.
When programming is complete
8.
Press Stop/Start.
9.
“Review pump settings” displays. Press Review.
10.
Choose Accept Value to confirm the value is correct
for the highlighted patient specific parameter or
press Select to edit the highlighted parameter.
11.
Continue until all patient specific parameters have
been reviewed, accepted and display checkmarks.
Press Next.
12.
“Start Pump?” displays. Press Yes.
Note: If a security code was used to unlock the keypad,
the pump will automatically relock when the pump is
started. If a key was used to unlock the cassette/keypad,
use the key to relock it.
13.
If you’re not starting the pump immediately, press
No when “Start Pump?” appears. Lock the keypad
by pressing the right soft key twice (Tasks then Lock
Keypad). Ensure that the cassette is also locked by
turning the cassette/keypad lock clockwise to the
locked position.
Resetting the reservoir volume without
changing the tubing