Smiths Medical CADD Solis v 3.0 Manual Download Page 6

Stop the pump

1. 

Press Stop/Start.

2. 

“Stop Pump?” displays. Press Yes.

3. 

Press the power switch to turn the pump off. 

Acknowledge “Power down?” by pressing Yes.

4. 

Remove the used batteries.

5. 

Insert the new batteries.

6. 

Press the power switch to turn the pump on.

7. 

The screen displays “Do you want to start a new 

patient?” Press No.

8. 

Press Stop/Start to start the pump.

9. 

“Start Pump?” displays. Press Yes.

Changing the batteries

Pump may be running or stopped to view reports.  

Pump reports should be cleared in accordance with 

institution policy.

Option 1:

1. 

From the home screen press Reports. Scroll 

 or 

 to the desired report and press Select.

2. 

Press Back to return to the Reports menu, and then 

press Back again to return to the home screen.

Viewing reports

Option 2:

1. 

From the home screen press Tasks. Scroll 

 to 

View Reports and press Select. Scroll 

 or 

 to the 

desired report and press Select.

2. 

Press Back to return to the Reports menu. Press 

Back again to return to the home screen.

To clear Given and PCA dose counters

1. 

From the home screen press Reports. Scroll 

 or 

 to the “Given and PCA Dose Counters” report. 

Press Select.

2. 

Press Clear Given to clear Total Given and update 

the date/time stamp.

3. 

Press Clear Doses to clear Given/Attempted and 

update the date/time stamp.

4. 

Press Back to return to the reports menu, and then 

press Back again to return to the home screen.

Summary of Contents for CADD Solis v 3.0

Page 1: ...ator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J Cassette Keypad Lock K Power Switch CADD Solis v 3 0 Ambulatory Infusion System with Programmed Intermittent Bolus Quick Reference Card for Clinicians ...

Page 2: ...en press Save Note If the desired value is outside the soft limit confirm the soft limit override by pressing Yes Note The next bolus setting allows for a one time override of the intermittent bolus cycle as defined by the bolus interval 13 Continue until all patient specific parameters have been reviewed and or edited Press Accept Value for each setting A check mark appears next to each patient s...

Page 3: ...t soft key twice Tasks then Lock Keypad If a key was used turn the key clockwise to relock the cassette and keypad 4 Verify that the keypad and cassette are locked Changing a patient s current program while the pump is running Pump must be running 1 From the home screen Press Tasks 2 Give Clinician Bolus is highlighted Press Select 3 Enter the clinician security code 4 The screen displays the clin...

Page 4: ... Review pump settings displays Press Review 8 Choose Accept Value to confirm the value is correct for the highlighted patient specific parameter or press Select to edit the highlighted parameter 9 Continue until all patient specific parameters have been reviewed accepted and display checkmarks Press Next 10 Start Pump displays Press Yes Note If a security code was used to unlock the keypad the key...

Page 5: ...iew pump settings displays Press Review 10 Choose Accept Value to confirm the value is correct for the highlighted patient specific parameter or press Select to edit the highlighted parameter 11 Continue until all patient specific parameters have been reviewed accepted and display checkmarks Press Next 12 Start Pump displays Press Yes Note If a security code was used to unlock the keypad the pump ...

Page 6: ... Scroll or to the desired report and press Select 2 Press Back to return to the Reports menu and then press Back again to return to the home screen Viewing reports Option 2 1 From the home screen press Tasks Scroll to View Reports and press Select Scroll or to the desired report and press Select 2 Press Back to return to the Reports menu Press Back again to return to the home screen To clear Given...

Page 7: ...y an amber screen it continues until acknowledged or until the condition that triggered the alarm goes away Low Priority Alarm A low priority alarm does not stop the pump Accompanied by a blue screen the alarm automatically clears after 5 seconds or when the condition that triggered the alarm goes away Informational Message This alarm does not stop the pump This message appears in the status bar I...

Page 8: ...en an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Charge or change the rechargeable battery pack or change the 4 AA batteries soon Current settings require high standard volume set Change cassette Alarm Priority High A high volume or standard volume administration set is required The pump is stopped and will not run Remove th...

Page 9: ...Reservoir volume is zero Pump stopped Alarm Priority High The reservoir volume has reached 0 0 ml The pump stops and can not run Acknowledge the alarm Install a new fluid container Reset or edit the value of the reservoir volume Upstream occlusion Clear occlusion between pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be resulting from a k...

Page 10: ...y business Smiths Group plc Please see the Instructions for Use Operator s Manual for a complete listing of the indications contraindications warnings and precautions CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2017 Smiths Medical All rights reserved Prod...

Reviews: