Smiths Medical CADD Solis v 3.0 Manual Download Page 2

Prepare the pump for a new patient

1. 

Begin without the cassette attached to the pump.

2. 

Insert four new 1.5 volt AA alkaline batteries or a 

rechargeable battery pack.

3. 

Press the power switch to turn the pump on.

Start New Patient

4. 

Screen displays “Do you want to start a new patient?” 

Press Yes.

5. 

The “Select Therapy” menu is displayed.

6. 

Scroll 

 or 

 to highlight the desired therapy.  

Press Select.

7. 

Scroll 

 or 

 to highlight the desired qualifier. 

Press Select.

8. 

Scroll 

 or 

 to highlight the desired drug and 

concentration (or unit). Press Select.

9. 

Unlock the keypad using the security code or  

the pump key.

10. 

Confirm that you have selected the correct therapy, 

qualifier, drug and concentration [or unit]. Verify and 

press Yes.

11. 

“Review pump settings” displays. Press Review.

12. 

To edit, highlight a patient specific parameter, press 

Select. Scroll 

 or 

 to the new value then press Save.

Note: If the desired value is outside the soft limit, confirm 

the soft limit override by pressing Yes. 

Note: The next bolus setting allows for a one time override 

of the intermittent bolus cycle as defined by the bolus 

interval.

13. 

Continue until all patient specific parameters have been 

reviewed and/or edited. Press Accept Value for each 

setting. A check mark appears next to each patient 

specific parameter you have accepted. To change a 

patient specific parameter after you have accepted it, 

repeat step 12.

14. 

When all values are accepted, press Next.

15. 

“Cassette not attached. Attach cassette before starting 

pump.” is displayed.

When programming for the new patient is complete

16. 

Attach, latch, and lock the cassette to the pump.

17. 

“Prime Tubing?” displays.  

Press Yes if priming is needed.

18. 

“Disconnect tubing…” displays. Press Prime.  

Press Stop Priming when complete.

19. 

“Continue Priming?” displays. Press Yes or No.

20. 

“Start pump?” displays. Press Yes when you are ready 

to begin the infusion. The pump begins running.

Setting up the Pump for a New Patient

Summary of Contents for CADD Solis v 3.0

Page 1: ...ator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J Cassette Keypad Lock K Power Switch CADD Solis v 3 0 Ambulatory Infusion System with Programmed Intermittent Bolus Quick Reference Card for Clinicians ...

Page 2: ...en press Save Note If the desired value is outside the soft limit confirm the soft limit override by pressing Yes Note The next bolus setting allows for a one time override of the intermittent bolus cycle as defined by the bolus interval 13 Continue until all patient specific parameters have been reviewed and or edited Press Accept Value for each setting A check mark appears next to each patient s...

Page 3: ...t soft key twice Tasks then Lock Keypad If a key was used turn the key clockwise to relock the cassette and keypad 4 Verify that the keypad and cassette are locked Changing a patient s current program while the pump is running Pump must be running 1 From the home screen Press Tasks 2 Give Clinician Bolus is highlighted Press Select 3 Enter the clinician security code 4 The screen displays the clin...

Page 4: ... Review pump settings displays Press Review 8 Choose Accept Value to confirm the value is correct for the highlighted patient specific parameter or press Select to edit the highlighted parameter 9 Continue until all patient specific parameters have been reviewed accepted and display checkmarks Press Next 10 Start Pump displays Press Yes Note If a security code was used to unlock the keypad the key...

Page 5: ...iew pump settings displays Press Review 10 Choose Accept Value to confirm the value is correct for the highlighted patient specific parameter or press Select to edit the highlighted parameter 11 Continue until all patient specific parameters have been reviewed accepted and display checkmarks Press Next 12 Start Pump displays Press Yes Note If a security code was used to unlock the keypad the pump ...

Page 6: ... Scroll or to the desired report and press Select 2 Press Back to return to the Reports menu and then press Back again to return to the home screen Viewing reports Option 2 1 From the home screen press Tasks Scroll to View Reports and press Select Scroll or to the desired report and press Select 2 Press Back to return to the Reports menu Press Back again to return to the home screen To clear Given...

Page 7: ...y an amber screen it continues until acknowledged or until the condition that triggered the alarm goes away Low Priority Alarm A low priority alarm does not stop the pump Accompanied by a blue screen the alarm automatically clears after 5 seconds or when the condition that triggered the alarm goes away Informational Message This alarm does not stop the pump This message appears in the status bar I...

Page 8: ...en an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Charge or change the rechargeable battery pack or change the 4 AA batteries soon Current settings require high standard volume set Change cassette Alarm Priority High A high volume or standard volume administration set is required The pump is stopped and will not run Remove th...

Page 9: ...Reservoir volume is zero Pump stopped Alarm Priority High The reservoir volume has reached 0 0 ml The pump stops and can not run Acknowledge the alarm Install a new fluid container Reset or edit the value of the reservoir volume Upstream occlusion Clear occlusion between pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be resulting from a k...

Page 10: ...y business Smiths Group plc Please see the Instructions for Use Operator s Manual for a complete listing of the indications contraindications warnings and precautions CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2017 Smiths Medical All rights reserved Prod...

Reviews: