- 7 - 

 

6.

 

Packaging and Deliverables 

 

6-1. Packaging 

One Liquid Tank and Controller are packaged into one box for delivery. 

The product is packaged with a vinyl bag for delivery.   

Accessories such as a power supply cable and Operation Manual are packaged together. 

The pacakging box is made specific for the Thermo Electric Bath. When it is returned for 

maintenance or repair to our factory, use this box. 

The packaging has been evaluated in accordance with JIS Z0200 level 

  (10 to 20kg). 

 

6-2. Packaging content 

Thermo Electric Bath (Liquid Tank) 

1 pcs 

Thermo Electric Bath (Controller) 

1 pcs 

Signal 

cable   1 

pcs 

DC cable    

 

 

1 pcs 

Power supply cable 

 

  

pcs 

Operation Manual

 

 

       

  

1  pcs 

 
 

7.

 

 Shipment inspection 

 

In assembly process of the Thermo Electric Bath, the following inspections are performed 100% and the 

inspection record is attached to the product for delivery.   
 

Table 2 Shipment inspection items 

 

Item 

Content 

1. Appearance form 

Defect on appearance and shape 

Name plate 

2. Leakage test 

Cooling water circuit 

Tank 

3. Standard temp. test 

Deviation of temp. from reference thermometer 

4. Temp. stability test 

Stability of temp. at set value 

5. Cooling capacity test 

Time to cool the recirculating fluid 

6. Dielectric withstand test 

Dielectric withstand test between AC and PE 

Summary of Contents for HEBC002-WA10

Page 1: ...BC002 WA10 HEBC002 WB10 Model no Series Part no Return this Specification to SMC after stamping of signing in the column of Signature of reception This Specification is recognized to be accepted if no reply for 20 days from the date of issue of SMC Signature of reception SMC date stamp Prepared by PD Div 6 Appr d Chk d Prep d Sakama PD Div 6 Saika PD Div 6 Ono PD Div 6 ...

Page 2: ...Rev No Date Reason for revision Appr d Prep d Note Revision History ...

Page 3: ......

Page 4: ...ow to Order and Manufacturing Code Indication 3 5 Specifications 4 6 Packaging and Deliverables 6 7 Shipment inspection 6 8 Functions 7 9 Outline Dimensions 10 10 Performance Chart 11 11 Communication 12 12 Precautions on handling 13 13 Connectors 16 14 Flow sheet 19 Contents ...

Page 5: ... improper cleaning and drying of the product cannot be compensated 2 It is recommended that you prepare a back up product to minimize machine downtime during repairs and maintenance 3 The packing box in which the product came has been made specifically for this product Use this box to return the product to the SMC factory for repair or maintenance 4 Separately the following parts need replacement ...

Page 6: ...ll belong to SMC 2 Period 1 years after delivery to the place customer specified 3 Scope If the product fails within the warranty period and the fault lies with SMC it will be repaired at no cost 4 Out of scope The following situations are out of the scope of this warranty The failure was caused by operation different from that shown in the Operation Manual or operation over the SMC specified para...

Page 7: ... repair When repair of the product is needed contact the shop where the product was purchased When this is done the repair will then be performed under warranty by SMC No cost repair of the product is assured within the period and requirements mentioned above Failures which occur after the warranty period has passed will in principle be charged 4 How to Order and Manufacturing Code Indication 4 1 ...

Page 8: ...anufacturing can be found on the model number plate attached to this product Mfg month Mfg year SERIAL No K O 0001 JAN 2006 Mfg month O Jan U Jul P Feb V Aug Q Mar W Sep R Apr X Oct S May Y Nov T Jun Z Dec Serial No Mfg year K 2006 L 2007 M 2008 N 2009 ...

Page 9: ...ply AC100 to 240V Single phase 50 60Hz 4 to 2A Overcurrent protection Circuit protector acting as a power supply switch with rated current 10A Main functions Auto tuning Set value memory Temp upper lower deviation limit alarm Output cutoff alarm Communication RS 485 RS 232C Input operation Indication Membrane key sheet 7 segment LED Alarm output Temp upper lower deviation limit alarm Output cutoff...

Page 10: ...a trade mark of Solvay Solexis and Fluorinert TM is a trade mark of 3M Note 4 An appropriate range is from 3 to 5L min To prevent damage to the radiating system do not supply a flow over the maximum flow rate of 8L min Note 5 When the temperature is set high the liquid temperature inside of the liquid tank and the temperature inside of the thermostat could differ greatly depending on the heating m...

Page 11: ...Liquid Tank 1 pcs Thermo Electric Bath Controller 1 pcs Signal cable 1 pcs DC cable 1 pcs Power supply cable 1 pcs Operation Manual 1 pcs 7 Shipment inspection In assembly process of the Thermo Electric Bath the following inspections are performed 100 and the inspection record is attached to the product for delivery Table 2 Shipment inspection items Item Content 1 Appearance form Defect on appeara...

Page 12: ...al setting the alarm comes on immediately after the power supply is turned on when the temperature at that time deviates from the set temperature by an amount outside of the upper or lower deviation limit It is possible to change this setting so that the alarm will not come on during warm up to a temperature within the upper and lower deviation limit after the power supply is turned on Table3 Rela...

Page 13: ...s generated via relay contact to the pin for the output cutoff alarm of the alarm output connector This alarm cannot be reset unless the power supply is turned off once and restarted Problems that may cause this alarm include the following Overheating of Liquid Tank Thermostat starts operating Lowering of Controller output voltage Stopping of Controller fun rotation Table 4 Relay contact for outpu...

Page 14: ...ommunication protocol RS 485 The transmission cable length i e transmittable distance is 500m RS 485 enables one master computer to set and monitor operation conditions of up to 31 Thermo Electric Baths HEBC002 WB10 This product has a communication function conforming to communication protocol RS 232C The transmittable distance is 15m RS 232C enables one master computer to set and monitor the oper...

Page 15: ... 11 9 Outline Dimensions The detail can be found on specifications Fig 2 Liquid Tank Fig 3 Internal dimensions of Liquid Tank Fig 4 Controller ...

Page 16: ...Cooling water 10 Ambient temperature 25 Recirculating fluid GALDEN HT200 Recirculating fluid level 180 At recirculating fluid temp 25 Cooling water flow rate 3L min Shut out from outside with a lid styrofoam Cooling water 25 R 20 25 30 35 40 45 50 55 60 65 0 10 20 30 40 50 60 70 80 Time min Recirculating fluid temperature Cooling water 10 Cooling water 35 Ambient temperature 25 Recirculating fluid...

Page 17: ...le6 Communication HEBC002 WA10 HEBC002 WB10 Protocol RS 485 RS 232C Circuit type Half duplex Half duplex Communication type Asynchronous Asynchronous Communication speed BPS 1200 2400 4800 9600 19200 1200 2400 4800 9600 19200 Character code ASCII ASCII Interface 2 wire 3 wire Parity Without Odd Even Without Odd Even Start bit 1 bit 1bit Data length 7 8 bit 7 8 bit Stop bit 1 2 bit 1 2 bit BCC chec...

Page 18: ...ric or magnetic field that generates surges Environments generating electrical static discharge and conditions in which static electricity is applied to this product Environments generating strong high frequency radiation including radio frequency appliances such as mobile phones and tranceivers Environments generating strong vibrations and impacts Environments that may be damaged by lightning Env...

Page 19: ...culating fluid or the product may idle and break The recirculating fluid has effective range of levels Check that it reaches that range under operating conditions 3 The recirculating fluid may evaporate The product is unable to maintain performance at a significantly lower fluid level and this could cause breakage of the circulating pump Maintain the appropriate level at all times 4 Check for the ...

Page 20: ...not used for a long time remove the recirculating fluid and cooling water and turn off the main power supply to prepare for unexpected accidents 4 Disposing When disposing of this product be sure to use an industrial waste processing vendor that conforms to the Law concerning disposal of waste and cleaning and the Ordinance defined by each municipal corporation Dispose of the recirculating fluid i...

Page 21: ...aboshi Electric Mfg co Ltd NJC 245 RM UL CSA Hirose Electric co Ltd CDA 15P Fixed screw M2 6 Connected Connected Connection cable DC cable Signal cable Nanaboshi Electric Mfg co Ltd NJC 245 PF UL CSA Hirose Electric co Ltd CDA 15S Fixed screw M2 6 Female Male Female Male Nanaboshi Electric Mfg co Ltd NJC 245 PM UL CSA Hirose Electric co Ltd CDA 15P Fixed screw M2 6 Connected Connected Controller c...

Page 22: ...pply cable Controller connector Power supply connector IEC60320 C 14 equivalent Male Connected Power supply cable Controller side IEC60320 C 13 equivalent Female AWG14 Content Black 1 AC100 240V L Black 2 AC100 240V N Green Yellow PE ...

Page 23: ...2 Temp upper lower deviation limit alarm common 3 4 Unused 5 Output cutoff alarm contact opened for alarm 6 Output cutoff alarm common 7 9 Unused Communication connector Hirose Electric co Ltd CDE 9S Fixed screw M2 6 Mating connector CDE 9P equivalent Content Pin no HEBC002 WA10 HEBC002 WB10 1 RS 485 T R A Unused 2 RS 485 T R B RS 232C RX 3 Unused RS 232C TX 4 Unused Unused 5 Unused RS 232C SG 6 9...

Page 24: ...d Tank Recirculating pump Cooling water inlet Cooling water outlet Tank Thermo module Temperature sensor Thermostat AC100 to 240V Alarm output connector Controller Signal connector DC output connector Communication connector RS 485 or RS 232C ...

Reviews: