Configuration
3.13
Figure 3.8 – Transducer Block – Transducer Type Selection
If is required an adjustment of unit, please just select the desired unit using the measured type
parameter according to the application. If the adjust requires a changing of measured values,
please, calibrate the device with reference, according to these steps:
•
Wait the process stabilizes and collect a sample;
•
Determine in laboratory the value of density/concentration of stabilized process.
Using the
Syscon
(or any configuration tool), the user can select the measured type and the lower
and upper calibration procedure.
If the user selects lower or upper page, the following window is shown and the user can see the
actual calibrated point, the primary value and status and the result of calibration procedure:
Figure 3.9 – Concentration / Density Calibration
The calibrated point must be inside of the sensor range limits allowed for each type of
concentration/density measuring.
The user can see
the actual
concentration/
density lower
calibrated point and
the matched
pressure value.
Summary of Contents for DT302
Page 7: ...Kpuvcnncvkqp Hnqyejctv XKK Kpuvcnncvkqp Hnqyejctv...
Page 8: ...FV524 Qrgtcvkqp Ockpvgpcpeg cpf Kpuvtwevkqpu Ocpwcn XKKK...
Page 34: ...DT302 Operation Maintenance and Instructions Manual 2 4...
Page 65: ...Maintenance Procedures 4 7 Figure 4 4 DT302 Exploded View Industrial Model...
Page 68: ...DT302 Operation Maintenance and Instructions Manual 4 10...
Page 79: ...Egtvkhkecvkqpu Kphqtocvkqp C07 EXAM and NEMKO...
Page 80: ...FV524 Qrgtcvkqp Ockpvgpcpeg cpf Kpuvtwevkqpu Ocpwcn C08 CEPEL...
Page 81: ...Egtvkhkecvkqpu Kphqtocvkqp C09 Control Drawing Factory Mutual FM...
Page 82: ...FV524 Qrgtcvkqp Ockpvgpcpeg cpf Kpuvtwevkqpu Ocpwcn C0...
Page 84: ...DT302 Operation Maintenance and Instructions Manual B 2...