Wait 4 to 8
minutes
4
How to Use
21
20
Mix
Wait while mixing
Figure M
4.1
Wait 4 to 8 minutes for the
auto-injector to mix your medicine.
The progress bar will gradually light
up, and you will hear steady ticking
during mixing (see Figure M).
•
Make sure the auto-injector is
standing upright on a flat surface. If
the auto-injector is not upright, the
medicine mixing will pause.
4.2
The auto-injector has finished the
automatic part of the mixing when
you hear
2
loud beeps and the entire
progress bar flashes.
Wait 4 to 8
minutes for the
auto-injector to
mix medicine.
Watch progress
bar gradually
light up.
Wait until you
hear 2 loud
beeps and the
entire progress
bar flashes.
4.3
Continue with Step 5 (page 22)
immediately after automatic mixing is
completed.
If you wait for more than
2
hours
before completing the steps for
mixing by hand (Step 5), the
auto-injector will automatically cancel
the procedure. If this happens, the
cartridge will be released and cannot
be used. To remove the cartridge,
see Step 10 (page 32).
•
If you still need to inject after the
auto-injector has canceled the
procedure, go back to Step 1 and
use a new cartridge (page 14).
If you see a slowly flashing green mixing
icon and the progress bar is frozen
b
page 42.
If you see a flashing red mixing icon
b
page 46.