4. SPECIFICATIONS
PAR Quantum
PAR Special
PAR Energy
Pyranometer
Lux
Working Range (1)
0-3000
µmol m
-2
s
-1
0-600
W m
-2
0-1500
W m
-2
0-150
klx
Sensor Passband
PAR
400 - 700 nm
350 - 1100 nm
CIE photopic
Curve V(
λ)
Housing
IP65
Enclosure: Light Grey ABS, PC (UV stabilised)
Cable Glands: 2* Grey Nylon
Dimensions
80mm x 58mm x 121mm (DxHxW)
including cable glands and feet
Weight
300g (with 3m cable)
Cable
7-1-4C military specification
Voltage Output
Customer choice of 0-1V up to 0-10V
Linearity error over
working range
<0.2%
Absolute calibration
error
Typically 3%, but <5%
Temperature
Co-efficient
± 0.1%/°C
Long term stability
(2)
± 2%
Response Time (3)
<10ms
Operating Range
-20°C to +70° C
0-100% RH
Power Supply
Requirements
Minimum power supply required = max 2V DC, up to 15 V DC
NOTES ON SPECIFICATIONS
(1) All Skye sensors will work at levels of irradiance well above that found in terrestrial sunlight conditions,
room or growth chamber lighting. These are default working ranges, check individual calibration certificate.
(2) Maximum change in one year. Calibration check recommended at least every two years. Experience has
shown that changes are typically much less than figures quoted.
(3) Times are generally less than the figure quoted, which is in milliseconds. They may be slightly increased
if long leads are fitted, or those of a higher capacity cable.