1 Channel External Voltage Amplifier

 

SKP 220V

 

Skye Instruments Ltd.,

 

21 Ddole Enterprise Park,

 

Llandrindod Wells,

 

Powys LD1 6DF UK

 

Tel: +44 (0) 1597 824811

 

   [email protected] 

 

www.skyeinstruments.com

 

 
 

Summary of Contents for SKP 220V

Page 1: ...85 v1a 1 Channel External Voltage Amplifier SKP 220V Skye Instruments Ltd 21 Ddole Enterprise Park Llandrindod Wells Powys LD1 6DF UK Tel 44 0 1597 824811 skyemail skyeinstruments com www skyeinstrume...

Page 2: ...ee to contact us via our e mail or any of the methods below Please click on the icons to browse the sites or search for the usernames Have a Smartphone Scan this QR code to access our website for more...

Page 3: ...1 Introduction 1 2 Wiring Power Details 2 2 1 Wiring 2 2 2 Connecting a Sensor 2 2 3 Sensor Output Power Supply 3 2 4 Connecting to Terminal 4 3 Housing 5 4 Specifications 6 5 Troubleshooting 7 CONTE...

Page 4: ...nge of amplified sensors have the amplification built into the sensor housing which is the ideal situation This is a product for those situations where this is not appropriate for example where there...

Page 5: ...the sensor cable through the cable gland and when in place finger tighten securely Be sure there is enough cable inside the box to reach the corresponding terminals The blue wire from the sensor must...

Page 6: ...ummer s Day PAR Special 0 3 000 mol m 2 s 1 2 000 mol m 2 s 1 PAR Quantum 0 3 000 mol m 2 s 1 2 000 mol m 2 s 1 Lux 0 150 000 Lux 100 000 Lux PAR Energy 0 600 W m 2 500 W m 2 Pyranometer 0 1 500 W m 2...

Page 7: ...etal clip inside Use a small flat bladed screwdriver supplied to press down firmly on the metal clip and insert the exposed wire into the terminal access located on the side of the terminal block Once...

Page 8: ...protected from ingress to IP65 On no account should it be subjected to immersion Dimensions of the amplifier unit are approx 80mm x 58mm x 121mm DxHxW this includes the cable glands and feet Please en...

Page 9: ...re Co efficient 0 1 C Long term stability 2 2 Response Time 3 10ms Operating Range 20 C to 70 C 0 100 RH Power Supply Requirements Minimum power supply required max output 2V DC up to 15 V DC NOTES ON...

Page 10: ...his can be an easily overlooked when encountering problems Also check that the correct wires are connected to the correct terminals Another reason that may account for this is inadequate power supply...

Page 11: ...Skye Instruments Ltd 21 Ddole Enterprise Park Llandrindod Wells Powys LD1 6DF UK Tel 44 0 1597 824811 skyemail skyeinstruments com www skyeinstruments com...