35
Cables and Accessories
Specifications and Requirements
Hand Control
Length: 11.16 feet (3.4 meters)
105°C (221°F)
24 AWG Strand
PVC Insulated
Standards: UL AWM Style 2464, NEC Type CMG/cUL CEC
Type CMG FT4, CSA AWM I A, EU CE
Power Cord
Hospital Grade Power Reel
Length: 10 feet (3.04 meters)
SJT 16/3C 60°C (140°F)
End Types: YP-18L-2 and YC-12
UL and CSA certified 3 prong IEC C13 power cord receptacle.
Built to the IEC 60320-C13 standard. Rated 10A 125V.
This IEC C13 power cord receptacle is fully molded with a low
profile ergonomic design and fully RoHS and REACH
compliant.
Electromagnetic Interference (EMI)
The Shuttle has been tested and is intended for safe use with other components compliant to IEC
601 standards for medical devices. The Shuttle can cause interference with non-601 regulated
equipment and/or other non-601 regulated equipment can cause interference with the
Shuttle. This section is in accordance with IEC 60601-1-2 section 5.2.2.
WARNING:
Radio wave sources such as radio stations, TV stations, amateur
radio (HAM) transmitters, two-way radios and cellular phones can affect this
Shuttle. Following the warnings listed for electromagnetic interference should
reduce the chance of the Shuttle’s unintended movement that could result in
serious injury.
DO NOT turn ON hand-held personal communication devices, such as citizens band (CB) radios
and cellular phones while the Shuttle is turned ON.
Be aware of nearby transmitters, such as radio or TV stations, and try to avoid coming close to
them.
If unintended movement occurs, turn the Shuttle OFF as soon as it is safe.
Adding accessories, components or modifying this Shuttle may make it more susceptible to
interference from radio wave sources (there is no easy way to evaluate their effect on the overall
immunity of the Shuttle).
Report all incidences of unintended movement to the Shuttle’s manufacturer and note whether
there is a radio wave source nearby.
CAUTION:
It is very important to read the information regarding the possible
effects of electromagnetic interference on the Shuttle.