![Sizewise SW Shuttle User Manual Download Page 40](http://html.mh-extra.com/html/sizewise/sw-shuttle/sw-shuttle_user-manual_1286088040.webp)
39
Table 206 Recommended Separation Distances between portable and mobile RG
Communications equipment and the Shuttle.
Equipment and Systems that are NOT Life-Supporting
Recommended Separations Distances for the Shuttle: The Shuttle is intended for use in the
electromagnetic environment in which radiated disturbances are controlled. The customer, or
user, of the Shuttle can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF Communications Equipment and the Shuttle, as
recommended below, according to the maximum output power of the communications
equipment.
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Compliance Level Cond RF 3 Rad RF-800 MHz 3 Rad RF – 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
General Safety
WARNING:
The Shuttle is to be used in accordance with each facility’s policies
and procedures.
CAUTION:
Possible fire hazard when used with oxygen administering
equipment other than the nasal, mask or 1/2 bed length type tent. Oxygen tent
should not extend below the mattress support level.
WARNING:
Electric shock may occur when plugging the Shuttle into the wall
outlet. Use ONLY grounded or hospital grade outlets.
DO NOT use the Shuttle if the power cord is cut, frayed or loosely connected to
the Shuttle.