23
Leg Supports
The leg supports are not designed to bear the full weight of the patient.
WARNING:
Never try to move the device by the leg supports. They may come
loose or break.
Never sit on leg supports when in the stretcher position.
DO NOT confuse head section and leg support areas of the
Shuttle when transferring patient in stretcher position.
DO NOT use the footrest as a step when the user sits on or
gets off the Shuttle.
To raise or lower, pull lever to elevate and push lever to release.
To extend or retract, push to release lever and adjust to the desired position.
To remove, extend leg support fully. Press the safety locking pin on the outside of the leg
support above the extension release lever.
Power System and Recharging
The Shuttle needs to be charged on a regular basis. The Shuttle is also designed to operate from
115V to 220V AC power source such as a standard wall outlet. The batteries are sealed and
service free. Do not exceed 6 hours of charge time.
AC Power
During initial setup of the Shuttle, it is recommended to charge batteries for a minimum 2-3
hours prior to use.
To connect the Shuttle to a wall outlet, use the retractable power cord provided on the left hand
side of the Shuttle. When the unit is connected, switch the unit power switch on the fuse box
located at the back of the Shuttle. A green light will indicate the Shuttle is energized. If the
Shuttle is connected to AC power source and the green light does not turn on, check the two in
line fuses. If needed, replace fuses as marked. Batteries are being charged while Shuttle is
plugged into wall outlet. DO NOT leave the Shuttle unattended when charging for more than six
hours.
Lever to raise
or lower leg
supports
Lever to extend
or retract leg
supports
Safety Lock
Push Pin