Safety Tips
Electrical Leakage Test Point
CAUTION:
If the ground is not properly connected to the electrical leakage test
point a false reading may occur.
In the event the electrical leakage test is done, the
proper location for this test is on the mounting bolt
for the cord reel.
Electromagnetic Compatibility (EMC)
The Shuttle has been tested for compliance with the EMC requirements. The guidelines in this
section will help to ensure the medical equipment will meet the requirements of the standard.
WARNING:
Medical equipment should not be used, stacked or located on or
around equipment that may create electromagnetic interferences.
Emissions
The Shuttle has been type tested and has passed the requirements of CISPR 11. Observe the
following recommendations to minimize radio frequency emissions in this section and the
Electromagnetic Interference section.
Immunity
The Shuttle has been stringently tested for susceptibility to electromagnetic radiation over the
frequency range 80 MHz to 2.5 GHz. The test was conducted on this Shuttle and passed the
requirements of IEC 61000-4-3.
All pins of connectors have passed ESD testing.
List of Cables and Accessories
Replacement parts, such as cables and accessories, must be purchased through Sizewise to
ensure proper compliance requirements.
WARNING:
Using other manufacturer’s cables and accessories may affect EMC
performance. Unauthorized use of these items will void warranty and could result
in patient and/or user injury as well as damage to equipment or other property.
The use of cables or accessories other than those for which the Shuttle was
designed or tested can significantly degrade emissions and immunity
performance.