Imaging
6-10
sweeping speed. Four speeds are available, including
1.25
,
2.5
,
5.0
, and
10.0
(second/frame).
6.4.9 IP One-key-optimization Function
When the current state is
Focus
,
M LINE
or
Wcont
in unfrozen state, press
I
to load the
first group of optimal parameters (which is prompted over the imagine area) and the image
changes accordingly. Press
I
again to load the second group of optimal parameters. Up to
8 groups of optimal parameters are available for selection.
The current system parameters can be saved to optimal parameters by pressing
U
key.
Press
U
key to save the current system parameters to that group of optimal parameters
loaded previously by pressing
I
key, (
Save Param
*
will be prompted over the imagine
area). If no
I
key was pressed previously, the parameters will be saved to the first group of
optimal parameters; press
U
again to save the current system parameters to the next
group of optimal parameters. Up to 8 groups of optimal parameters are available for
update. To save the current system parameters to a desired group of optimal parameters,
press
I
key first to select and load to the desired group of optimal parameters and press
U
to save.
The optimal parameters can be restored to default values by pressing
Y
key. Press
Y
key
to restore the group of optimal parameters previously loaded by pressing
I
key, which is
prompted over the imagine area. If no
I
key was pressed previously, it will default the first
group of predefined optimal parameters; press
Y
again to default the next group of
predefined optimal parameters. Up to 8 groups of optimal parameters are available for
restore. To restore a desired group of optimal parameters to default values, press
I
key to
select and load the desired group of optimal parameters and press
Y
key to restore
default values.
6.4.10 Adjusting Power
In unfrozen state, press
T-Ball
key and use the touchpad to select
PWR
in the control
window. Press
Value
control key to control the acoustic power strength. The acoustic
output parameter changes as the acoustic power strength is adjusted.
【
Warning
】
: To avoid harm to the patient, it is recommended to use as low acoustic
power as possible provided that a clear image is obtained.