Installation, Inspection and Maintenance
4-3
different capacity, model or type;
o) If the battery is leakage or gives off peculiar smelling, move the battery away
from open flame, because leaked electrolyte may cause fire or explode.
p) Do not use the battery if it is in peculiar smelling, heats, distortion, changed
color or any other abnormity. If the abnormal battery is charging or working,
unload the battery from the system or charger, and stop using.
q) Do not wipe but use water to wash away immediately and go to the hospital for
help in time, if any electrolyte due to battery leakage goes into eyes by
accident.
Note 1: When the system is used for the first time, please take out the film in
the battery compartment.
Note 2: A battery capacity indicator is displayed on the screen. When the
battery is full, the battery indicator occupies three boxes. With the
consumption of power (the battery indication capacity decreases
gradually), the battery boxes occupied decrease. When the battery
capacity is lower than the set value, the power indicator will flash. Now
you need to plug in the power supply adapter immediately to charge
the battery, or shut off the system to replace the battery.
Note 3: If the battery is not in use for a long period, please take it out and store
it in a cool place. Do not dispose the battery into fire, otherwise it may
cause explosion.
4.1.2 Components Connection
a) Press the inner side of the buckle on the probe connector to make it open. Plug
the probe connector right into the probe socket on the main unit (Please make
sure the assembly orientation of the connector). Loosen the buckle when make
sure the connector is well connected with the main unit. Follow the above steps
reversely when taking out the probe connector;
b) If a video printer shall be connected, connect its VIDEO IN port with VIDEO OUT
port on the main unit via a BNC video cable, and connect its REMOTE port with
PRINTER REMOTE port on the main unit via a print control cable;
c) If a USB printer shall be connected, connect its USB port with the USB port on the
main unit;
d) If images shall be transferred via a DICOM 3.0 port, connect a network cable to
the network port RJ-45 on the main system;
e) When connecting the DC power plug to the main unit, please align the raised