Acoustic Output Parameter
D-2
Transducer Model
:
C3I60H
Operating
Mode
:
B
TIS TIB
Non-scan
Index Label
MI
Scan
A
aprt
≤1cm²
A
aprt
>1cm²
Non-
scan
TIC
Global Maximum index value
0.845
0.278
0.635
P
r
α
Mpa 1.523
P
mW
42
42
Min.of[
P
α
(
z
s
)
,
I
ta,
α
(
z
s
)
]
mW
z
s
cm
z
bp
cm
z
b
cm
z
at max.
I
pi,
α
cm 4.05
d
eq
(z
b)
cm
f
awf
MHz 3.244 3.244
3.244
X
cm
1.92
Associated
Acoustic
parameters
Dim of
A
aprt
Y
cm
1.12
t
d
μ
sec 0.602
p
rr
Hz 8000
p
r
at max.
I
pi
Mpa 2.854
d
eq
at max.
I
pi
cm
Other
information
I
pi,
α
at max.
MI
W/cm²
247.76
Control 1:
Power
dB
0
0
0
Control 2:
Focal Depth
cm 6.0 6.0
6.0
Control 3:
Scan Angel
。
70 70
70
Operating
Control
conditions
Control 4:
Frequency
MHz 3.5 3.5
3.5