Measurement and Calculation
8-28
displayed in the measuring result area.
Fig. 8-13 Initial, Final and Result Page of TV measurement
8.3.10.4 PV measurement
Fig. 8-14 Initial, Final and Result Page of PV measurement
PV is Pulmonic Valve measurement in frozen state in M or B/M mode. Enter PV
measurement. On the left of the image there is a PV test module. A
+
cursor appears in
the screen center. Move the cursor with the touchpad to point
A
. Press
Enter
key to
input point
A
and the
+
cursor is fixed. Points
B, C, D
and
F
will be shown. Operate the
touchpad to move the
+
cursor to point
B
and
E
in M mode in the same way. The PV
Meas/Calc result will be displayed in the measuring result area.
8.3.10.5 LV measurement
LV is Left Ventricle measurement in frozen state in M or B/M mode. Enter LV
measurement and an LV test module appears on the left of the image. A vertical dotted
line and a
+
cursor are displayed in the screen center. Move the cursor with the touchpad
to the position corresponding to point
C
. Press
Enter
key to input point
C
, and point
D
will be displayed. Operate the touchpad to move the
+
cursor to the point on M mode
image corresponding to point
D
, in the same way as that for point
C
. Enter other points in
order. When the
+
cursor is placed on corresponding point
L
, the LV Meas/Calc result will
be displayed in the measuring result area. As points
C
~
F
and
G
~
J
lie on the vertical dot
lines
A
and
B
, the cursor can be moved up or down (rather than left or right) along the