7 Care of outer surface
Sirona Dental Systems GmbH
7.1 Care and cleaning agents
Operating Instructions XIOS
Plus
Sensors
7
Care of outer surface
Care of outer surface
7.1
Care and cleaning agents
Care and cleaning agents
CAUTION: Approved care and cleaning agents
Care and cleaning agents, with REF number
A continuously updated list of approved agents can be downloaded from
the Internet at the address
"www.sirona.com". In the navigation bar, go to
the menu items
"SERVICE" / "Care and cleaning" and then open the
document
"Care and cleaning agents".
If you do not have any access to the Internet, please contact your dental
depot to order the list.
REF 59 70 905
Care and cleaning agents, supplement for USA
Care and cleaning agents, supplement for USA
The outside areas may be disinfected with a suitable chemical
disinfectant. Only use disinfectants that meet the requirements of the
national authorities and have been tested and certified for the necessary
bactericidal, fungicidal and virucidal properties.
Examples of approved disinfectants are:
● MinutenSpray classic, by ALPRO®
● MinutenWipes, by ALPRO®
In the USA and Canada:
● CaviCide® or
● CaviWipes ™ .
7.2
Disinfecting
Disinfecting
Overview
The following components can only be disinfected by wiping them off:
● Size 1 and size 2 XIOS
Plus
X-ray sensors
NOTICE
Approved care and cleaning agents
Use only care and cleaning agents which have been approved by
Sirona!
NOTICE
Risk of sensor destruction
The sensor plug contains sensitive electronics.
No fluids may penetrate the sensor plug during preliminary cleaning and
wipe disinfection!