Sirona Dental Systems GmbH
5
Use of the X-ray Sensor
Operating Instructions XIOS
Plus
Sensors
General information
båÖäáëÜ
5
Use of the X-ray Sensor
Use of the X-ray Sensor
General information
General information
Placement
The X-ray sensor should be placed according to the well-known parallel
or half-angle technique.
Risk of damage to sensor
XIOS Plus sensor operational reliability
5.1
Preparation
Preparation
Checklist
CAUTION
Illustrations without hygienic sleeves
For the sake of clarity, no hygienic sleeves are shown in some of the
illustrations. However, a hygienic sleeve always must be slipped over
the sensor before using it on a patient.
NOTICE
Risk of damage
- Handle the sensor carefully.
- Do not drop the sensor!
- Do not swirl the sensor about by its cable.
- Do not bend the sensor cables or run over them (e.g. with a chair)!
- Avoid laying sensor cables on the floor if possible.
- Make sure that the patient does not bite on the sensor or the sensor
cable.
NOTICE
Operational reliability
- The sensor must be checked to ensure that it is plugged in correctly
before releasing an exposure!
- The user must check exposure readiness prior to radiation release.
See Operating Instructions for the XIOS
Plus
wall module or the XIOS
Plus
USB module.
- Never place two sensors in the beam path at the same time!
NOTICE
Check the sensor cable and sensor for damage, e.g. cracks or splinters,
prior to each exposure. In case of visible damage to the sensor, please
contact your dealer.