02 08.2011
5 Use of the X-ray Sensor
Sirona Dental Systems GmbH
5.3 Parallel technique with radiation limiter!
Operating Instructions XIOS
Plus
Sensors
Preparation
1. Plug together the straight guide rod (C) and the red localizer ring (B)
of the XIOS holder system for bite wing exposures.
2. Select the (bite tab type) red sensor holder tab (A) for bite wing
exposures and plug it onto the guide rod (C).
3. Slide the sensor into the hygienic protective sleeve, When doing so,
follow the instructions for sensors. [ → 19]
4. For vertical bite wing exposures: Position the sensor holder tab
vertically on the hygienic protective sleeve in the middle of the active
area of the sensor as shown in the illustration. The active area of the
sensor is identified by dots on the sensor.
5. For horizontal bite wing exposures: Position the sensor holder tab
horizontally on the hygienic protective sleeve in the middle of the
active area of the sensor as shown in the illustration. The active area
of the sensor is identified by dots on the sensor.
X-ray image
1. Position the sensor in the patient's mouth.
2. Place the X-ray tube assembly in the correct position. Change the
position of the sensor holder if necessary.
3. Release an X-ray exposure.
4. Discard the sensor holder tab and the hygienic protective sleeve
following the patient examination.
5. The guide rod and localizer ring must be sterilized.
5.3.4
Endodontics
Endodontics
Explanation
The "endo" type sensor holder tab is available for endodontics.
● This sensor holder tab, the matching guide rod and the matching
localizer ring are color-coded gray.