63 74 198 D3545
D3545
.
201.06.03
.
02 05.2014
25
Sirona Dental Systems GmbH
Installation
Operating Instructions SIROLaser Advance
Initial start-up – procedure for proper assembly
båÖäáëÜ
4.4
Initial start-up – procedure for proper assembly
Any national or local regulations stipulating that the SIROLaser Advance
may be installed only by trained personnel must be strictly observed.
With foot control
1. Install power supply
2. Assemble handpiece, optical fiber and single-use tip
3. Install wireless foot control – optional
4. Install remote interlock – optional
5. Start SIROLaser Advance for the first time
4.4.1
Install power supply
1. Connect the power cable to the DC IN socket at the back of the
SIROLaser Advance.
2. Please make sure to switch on the switching power supply.
The green LED on the power supply lights up.
Power supply MPU100-106
With battery
The SIROLaser Advance is supplied with a rechargeable battery and
therefore can be used without connected power cable. The status of the
rechargeable battery and whether the power cable is actually connected
will be always displayed on the touch screen.
The SIROLaser Advance is fully functional and can be run while charging
the battery.
➢ Charge the battery completely.
CAUTION
The SIROLaser Advance may only be operated with the Sinpro
MPU100-106 power supply. Operation with other power supplies may
result in failure or destruction of the laser unit. If any power supply other
than the one recommended is used, the approval of the entire unit
automatically becomes void and the warranty granted by Sirona Dental
Systems GmbH expires.
The use of any power supplies other than the one recommended may
cause overheating and failure of the laser unit or damage of batteries.
NOTICE
There will be a warning if the rechargeable battery will reach a low level
of capacity.
NOTICE
The rechargeable battery must be fully charged regularly. After six
months of non-charging the rechargeable battery might reduce its
loading capacity.