Installation
Operating Instructions SIROLaser Advance
Initial start-up – procedure for proper assembly
båÖäáëÜ
8. If the aiming beam projects a uniform circular beam pattern, remove
the distal seal with the fibercutter, see chapter „Adjusting the optical
fiber with the fiber cutter“ [ ➙ 31]. Any liability of the manufacturer
regarding mechanical damage to the optical fiber is thereby
canceled.
9. Insert the handpiece body back into the handpiece sleeve making
sure that the optical fiber slides easily through the single-use tip. Also
check if the snap tabs have clicked into their rests.
10. Now you can insert the single-use tip and the optical fiber into the
bending tool and bend the single-use tip to the angle that you need
for best handling.
empty
After treatment
Please bend back the single-use tip as straight as possible before moving
the fiber within the single-use tip.
As soon as you disassemble the optical fiber after treatment make sure
to protect the optical fiber socket and the optical fiber with the protective
caps provided for this purpose.
For cleaning, disinfecting and sterilizing the optical fiber please refer to
chapter "Cleaning, disinfection and sterilization" [ ➙ 62].
2.
135C
°
1.
1.
2.
3.
CAUTION
Cover the optical fiber socket and the optical fiber with the special
protective caps provided for this purpose after every treatment. Prevent
dust, dirt and foreign particles from entering the optical fiber socket. Make
sure that the optical system is clean before connecting the optical fiber.
Make sure that no dust or dirt can enter the optical fiber socket or the
optical system. Otherwise the unit may be permanently damaged.
If the optical fiber is moved in a bent single-use tip the surface of the
optical fiber may be destroyed (danger of optical fiber breakage).