63 74 198 D3545
32
D3545
.
201.06.03
.
02 05.2014
Installation
Sirona Dental Systems GmbH
Initial start-up – procedure for proper assembly
Operating Instructions SIROLaser Advance
✔
The SIROLaser Advance control unit and the wireless foot control are
ready for operation.
1. Choose in main home screen the item "Set-up".
2. Choose there the item "Activation device".
3. Choose the "Wireless registration".
4. Follow the instructions in the note window and press first the foot
switch for three seconds.
5. After this press the registration key on the top of the wireless foot
control radio box for three seconds.
After release of the registration key the note window disappears.
This means the wireless foot control is activated.
6. To use the wireless foot control, choose in submenu "Set-up" in
"Activation device" the wireless foot control.
4.4.4
Install remote interlock – optional
empty
Explanation
The interlock is a safety device that stops laser radiation whenever the
door of the treatment room is opened. The interlock circuit must be
connected to a switch that is located near the door of the treatment room
in order to ensure automatic interruption of the laser emission.
empty
Installation of an interlock with door switch
1. Prepare the interlock plug by connecting the interlock cable with the
interlock plug and by opening the bridge. Please find the technical
data sheet with circuit diagram for the installation of the interlock
circuit in "Appendix C" [ ➙ 79].
2. Mount the prepared interlock plug into the interlock socket on the
backside of the SIROLaser Advance.
NOTICE
The finger switch is pre-set.
NOTICE
The installation must be performed by a qualified electrician who is also
responsible for the installation and maintenance of the electrical system
to which the SIROLaser Advance is connected.
NOTICE
Additional or different safety precautions required by the applicable
national or local regulations for the protection of dentists, assistant
personnel, or patients must also be observed.