Cleaning, disinfection and sterilization
Sirona Dental Systems GmbH
Cleaning
Operating Instructions SIROLaser Advance
7
Cleaning, disinfection and sterilization
Following treatment, switch off the SIROLaser Advance and disconnect
the power cable from the power supply.
With foot control
Control unit, handpiece body, handpiece tube and foot control are
supposed to be cleaned and wipe-disinfected.
Dispose the single-use tip.
The removable Handpiece sleeve, stainless steel back-sleeve, optical
fibers, fibercutter and bending tool are supposed to be cleaned and
sterilized.
For the number of sterilization cycles, see chapter "Replacement of parts
subject to wear and tear" [ ➙ 69].
7.1
Cleaning
empty
Optical fiber and handpiece sleeve
Before disassembly the optical fiber must be cleaned from residual tissue
and disinfected.
1. Please bend back the single-use tip as straight as possible before
moving the fiber within the single-use tip.
2. Notch the optical fiber approx. 4 mm from its distal end. The notch
must be made perpendicular to the axis of the optical fiber.
3. Remove the single-use tip from the handpiece and dispose it.
4. Remove the handpiece sleeve from the handpiece body by pressing
both snap tabs.
5. After removal of the optical fiber from the handpiece, the optical fiber
can be cleaned with a suitable cloth under running water (of at least
drinking water quality).
NOTICE
When following these instructions wear gloves.
CAUTION
Do not clean and disinfect the parts using a washer-disinfector! The
parts may be seriously damaged.
NOTICE
Manual cleaning must always be combined with disinfection.
All tissue residues must be removed from the optical fiber before it is
detached from the handpiece.
CAUTION
If the optical fiber is moved in a bent single-use tip the surface of the
optical fiber may be destroyed (danger of optical fiber breakage).