6.1.
Environmental conditions
The lifetime of t
he
electrochemical cartridges
for toxic gas detector depends on the
operating environment related to temperature, humidity and high exposures.
One will take care to avoid long exposures to a hot and dry atmosphere (T°> 30°C and
RH<10%), and gas exposures 4 times higher to the scale.
The lifetime of
the cartridges for semi-conductor toxic gas detector
depends on the
operating environment related to certain compounds.
One will take care to avoid exposures to some vapored products as silicone. Moreover, an
exposure in area with a low rate of hygrometry for several hours can temporarily damage the
sensitivity. In order to retain stability of the factory span calibration, we recommand to leave
in place the protection cap with a desiccant packet on the front of the cartridge during
storage, shipping or any periods without power lasting more than 1 hour.
Don’t forget to remove the protection cap after power ON and before inserting
the eventual splash-guard.
Store the protection caps with the desiccant packet in a sealed container (zip-
lock bag). Dry the bag before using it again
In general, a dusty and humid atmosphere must be avoided. Indeed, a clogging of the gas
barriers is then possible, slowing down or stopping the detection process.
Summary of Contents for MultiTox DG-TT7-S
Page 1: ...Operating Manual NOSP 16452 05 03 2018 MultiTox Detector MultiTox DGi TT7 E DGi TT7 O DG TT7 S...
Page 2: ...MultiTox DGi TT7 E O DG TT7 S Page 2 60 NOSP 16452 05 08 2016...
Page 15: ...MultiTox DGi TT7 E O DG TT7 S Page 15 60 NOSP 16452 05 08 2016 Figure 2 Outline drawing mm...
Page 39: ...MultiTox DGi TT7 E O DG TT7 S Page 39 60 NOSP 16452 05 08 2016...
Page 54: ...MultiTox DGi TT7 E O DG TT7 S Page 54 60 NOSP 16452 05 08 2016 13 DECLARATION OF CONFORMITY...
Page 55: ...MultiTox DGi TT7 E O DG TT7 S Page 55 60 NOSP 16452 05 08 2016...
Page 56: ...MultiTox DGi TT7 E O DG TT7 S Page 57 60 NOSP 16452 05 08 2016...
Page 57: ...MultiTox DGi TT7 E O DG TT7 S Page 58 60 NOSP 16452 05 08 2016...
Page 58: ...MultiTox DGi TT7 E O DG TT7 S Page 59 60 NOSP 16452 05 08 2016...
Page 59: ......