MultiTox : DGi-TT7-E/O & DG-TT7-S
Page 35 / 60
NOSP 16452-05 (08/2016)
6.7.
Information menu [INFO]
The information menu contains all information concerning the identity and settings of the
detector. The first screen gives the detector’s reference and its serial number.
6.7.1.
[IDEN]tity submenu
Presentation of:
•
The serial number
•
The device reference
•
The scale and the targeted gas
Sub-menus present the board software version, the power supply voltage and the
temperature read in the cartridge.
6.7.2.
[OUT]put submenu
Presentation of:
•
Current protocol (0-20 mA or 4-20 mA).
•
Normal state of the relays (normally open or normally closed).
•
Condition of relay activation.
Relays can be set with a level 2 access.
6.7.3.
[STAT]e Information submenu
Presentation of:
•
Number and value of activated alarm levels.
•
List of eventual faults (press F1 key to scroll through the list)
•
Possibility to acknowledge alarms
6.7.3.1.
[ALRM] Alarm screen
Allows setting of levels and alarms acknowledgement.
6.7.3.2.
[FLT.] Fault screen
Displays a list of eventual faults (press F1 key to scroll through the list) and allows their
acknowledgment.
6.7.3.3.
Network screen: switch state
This menu and its sub-menus are used for the network detector settings. For any further
details, please refer to the Syntel system operating manuals.
MODE SCREEN
The first line shows the operating mode of the sensor in the network (logic link test / out of
order / emulation).
The second line shows if the network part of the detector is “operating” or “out of order”. For
any further details, please refer to the Syntel system operating manuals.
NETWORK SCREEN: ALIM
Information displayed:
•
Voltage A: ON / OFF
•
Voltage B: ON / OFF
For any further details, please refer to the Syntel system operating manuals
Summary of Contents for MultiTox DG-TT7-S
Page 1: ...Operating Manual NOSP 16452 05 03 2018 MultiTox Detector MultiTox DGi TT7 E DGi TT7 O DG TT7 S...
Page 2: ...MultiTox DGi TT7 E O DG TT7 S Page 2 60 NOSP 16452 05 08 2016...
Page 15: ...MultiTox DGi TT7 E O DG TT7 S Page 15 60 NOSP 16452 05 08 2016 Figure 2 Outline drawing mm...
Page 39: ...MultiTox DGi TT7 E O DG TT7 S Page 39 60 NOSP 16452 05 08 2016...
Page 54: ...MultiTox DGi TT7 E O DG TT7 S Page 54 60 NOSP 16452 05 08 2016 13 DECLARATION OF CONFORMITY...
Page 55: ...MultiTox DGi TT7 E O DG TT7 S Page 55 60 NOSP 16452 05 08 2016...
Page 56: ...MultiTox DGi TT7 E O DG TT7 S Page 57 60 NOSP 16452 05 08 2016...
Page 57: ...MultiTox DGi TT7 E O DG TT7 S Page 58 60 NOSP 16452 05 08 2016...
Page 58: ...MultiTox DGi TT7 E O DG TT7 S Page 59 60 NOSP 16452 05 08 2016...
Page 59: ......