GD1
Page 38 of 76
850-816926-EN R06
The above spectrum is found in GD1 Web interface. Spectrum to the left showing
472 ppm of CO
2
in the air. To the right a spectrum showing 1 191 ppm as a result of
introducing the Test Cell filled with 40 000 ppm CO
2
.The value can be verified by using the
formula in step 4.
Note!
Depending on the precision of the gas mix inside the Test Cell, do not expect to see an
exact 1:1 response compared to the test gas concentration.
5
Run loop testing in the Web interface "I/O
verification" screen.
Recommend to use 60 sec test time.
Enter test value. For example 12 mA is 50% of full
scale.
6
After the GD1 goes into "IO Verify" mode, verify
that the analogue output matches the test value.
Functional test with CO
2
finished!
Summary of Contents for GD1 MK3
Page 1: ...P N 850 816926 EN Rev6 Operating Manual Simtronics GD1 MK3 Toxic Open Path Gas Detector...
Page 8: ...GD1 Page 8 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 20: ...GD1 Page 20 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 42: ...GD1 Page 42 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 54: ...GD1 Page 54 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 58: ...GD1 Page 58 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 70: ...GD1 Page 70 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 72: ...GD1 Page 72 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 73: ...GD1 Page 73 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 74: ...GD1 Page 74 of 76 850 816926 EN R06 THIS PAGE LEFT INTENTIONALLY BLANK...
Page 75: ...GD1 Page 75 of 76 850 816926 EN R06 EU DECLARATION OF CONFORMITY...