![SIMTEC TRACSCAN 2.0 Operating Manual Download Page 26](http://html1.mh-extra.com/html/simtec/tracscan-2-0/tracscan-2-0_operating-manual_1279387026.webp)
Operating Manual TRACSCAN 2.0
V2.0, 2022-05
26 / 36
Error
Description
How to fix
Invalid License
Key
Entered License Key is not valid.
Contact service.
License Expired
License has expired and must be
renewed.
Contact service.
Limited Operat-
ing Period
Instrument is for trial use, only.
Operating Period will expire soon. Contact service.
License Expired
Instrument is for trial use, only.
Operating Period has expired.
Contact service.
Not Calibrated
Instrument has not been cali-
brated.
Contact service.
Calibration Re-
quired
Calibration Period has expired. In-
strument must be re-calibrated.
Contact service.
Calibration Due
Calibration Period will expire in 30
days or less.
Contact service.
Test Foot Missing Test Foot missing or not detected.
Check if a test foot is plugged at
all. Check if the test foot is plugged
properly. Make sure that the con-
tact pads are clean.
Test Foot Data In-
valid
Test Foot has been detected but
data could not be read or written.
Check if the test foot is plugged
properly. Make sure that the con-
tact pads are clean.
Cannot Lift Sen-
sor
Test foot cannot be lifted.
Contact service.
Cannot Lower
Sensor
Test foot cannot be lowered.
Contact service.
Invalid Request
Attempt to start unknown pro-
gram.
Contact service.
Invalid Program
Settings
One or several parameters of the
selected program are invalid.
Make a printout of all settings and
contact service.
Sample Data
Overflow
The driving distance or time is
longer as estimated.
Same as for Wheelspin.
Sample Data Un-
derflow
The measurement distance (DCOF
or TCOF) is too short.
Check program settings.
Strain Gauge
Overload
The strain gauge is in overload,
i.e., the traction force is too high. Contact service.
Strain Gauge
Fault
The strain gauge is broken or not
connected; the traction force can-
not be measured.
Contact service.