![SIMTEC TRACSCAN 2.0 Operating Manual Download Page 10](http://html1.mh-extra.com/html/simtec/tracscan-2-0/tracscan-2-0_operating-manual_1279387010.webp)
Operating Manual TRACSCAN 2.0
V2.0, 2022-05
10 / 36
4.2.
Licensing
If the device is only available on a trial basis, measurements cannot be taken if the period of
operation expires. Contact your distributor to obtain a license key for unlimited operation.
4.3.
Charging
The TRACSCAN 2.0 is fitted with a rechargeable battery. A standard USB A/B cable and charger
is included for recharging the battery, see section [9] for technical data.
The charging time with the included charger is up to 12 hours. The instrument has an overload
protection, but it is best not to leave the charger plugged in after the light has turned solid
green (1 to 2 days maximum). This will extend the battery life.
Any standard USB power bank can be connected even during operation, both charging the
battery and extending the operation time.
The LED next to the battery symbol on the control panel shows the charging state:
Color
Meaning
Solid Green: Battery fully charged
Blinking Green: Charging
Solid Red: Battery or charging fault, see section [7]
Table 1: Battery Charging LED
4.4.
Printing
The instrument is equipped with a thermal printer. It can be used to print out all instrument
and program parameters as well as the measurement results on an individual basis.
What to Print
Short Report of most recent
measurement
Parameter [Function of Print Key] must be set to “Short Re-
port”.
Press Print Button if instrument is off or right after meas-
urement has been taken.
Long Report of most recent
measurement
Parameter [Function of Print Key] must be set to “Long Re-
port”.
Press Print Button if instrument is off or right after meas-
urement has been taken.
Short Report of any archived
measurement
Open Instrument Menu and navigate to
Archive … / Measurements …
Select desired record and then
Short Report
Long Report of any archived
measurement
Open Instrument Menu and navigate to
Archive … / Measurements …
Select desired record and then
Long Report