4.2 AeroBar Maintenance
Primary maintenance for the AeroBar consists of periodic emitter
point cleaning and replacement, system calibration, and exterior
cleaning of the chassis. As maintenance schedules will vary
depending on conditions, develop a schedule which meets the
requirements for your application. In general, equipment should be
checked on a monthly basis to ensure it is operating as originally
set.
Emitter point cleaning is recommended every three to six months,
and replacement every 24 to 36 months or when damage or erosion
is evident. Calibration may be performed quarterly, semiannually, or
annually depending on your application. Always clean the emitter
points prior to calibration. Cleaning emitter points after calibration
will alter the calibrated settings.
Where AeroBars are used in environments containing airborne
molecular contaminates (AMC), emitter points should be cleaned
every one to three months or wherever significant amounts of debris
have accumulated on emitter tips.
Figure 9. Clean and Dirty Silicon Tips
Clean, new silicon tip
Eroded and dirty silicon tip
Summary of Contents for AeroBar 5225S
Page 5: ...19 5225S M 01 Rev 2 1 1 Description 1 1 Overview 1 2 Remote Control 1 3 Power Requirements...
Page 12: ...19 5225S M 01 Rev 2 8...
Page 42: ...19 5225S M 01 Rev 2 38...
Page 51: ...19 5225S M 01 Rev 2 47...
Page 57: ...19 5225S M 01 Rev 2 53 5 3 Dimensional Drawings Standard Lengths...
Page 58: ...19 5225S M 01 Rev 2 54 300 mm EFEM Application Lengths...
Page 59: ...19 5225S M 01 Rev 2 55...
Page 61: ...19 5225S M 01 Rev 2 57...
Page 62: ...19 5225S M 01 Rev 2 58 Notes...