Mounting
20 / 176
SIGRIST-PHOTOMETER AG
2.4
Specification sheet
Photometer
Values
Measuring principle
90° / 25° scattered light measurement at 650 nm (optional colour mea-
surement at 430 nm)
Explosion protection type / tempera-
ture class
Ex db IIC T3/T4/T5/T6 Ga/Gb, temperature class depending on medium
temperature:
-20 … +80 °C = T6
-20 … +95 °C = T5
-20 … +130 °C = T4
-20 … +180 °C = T3
Ambient humidity
0 ... 100 % relative humidity
Ambient temperature
-20 … +60 °C
Medium temperature
Cooling depends on the maximum medium temperature (Tmed.) as well
as on the ambient temperature (Tenv.). The shaded area
(X)
indicates the
temperatures above which cooling by means of an optional cooling ring is
required. The flow must be at least 0.2 l/min. at a coolant temperature of
max. 20 °C.
X
Max. pressure
TurBiScat 2 Ex (window insert): 2 MPa (20 bar) at max. 180 °C.
VARINLINE
®
connector with blanking plate (art. no.: 122037): Note speci-
fication.
Flange connection with special measuring cell: 2 MPa (20 bar) at max.
180 °C.
Service voltage
24 VDC ± 10 %
Protection class
IP 66
3
Mounting
3.1
Requirements
For the mounting of the photometer and its additional components, the detailed dimension drawings must be ob-
served. The following provisions apply to the applications.
The photometer must be installed in the line at least 2 m away from sight glass or other sources of interfering
light.
Summary of Contents for TurBiScat 2 Ex
Page 88: ...88 176 SIGRIST PHOTOMETER AG l l l l l l l l l l l l l 1 4 u 1 5 1 6...
Page 94: ...94 176 SIGRIST PHOTOMETER AG 4 3 0 2 mm2 100 m 10 5 TOUCH WiFi 5 1 l 0 5s l 1 2s 5 cm 6 u u...
Page 99: ...SIGRIST PHOTOMETER AG 99 176 u u u RMA u u u u u RMA RMA DE EN FR ES NL PT ZH RU CS JA IT PL...
Page 112: ...112 176 SIGRIST PHOTOMETER AG u u 5 u 4 3 u 3 u 1 u 6 u 7 u u 2 1 3 4 5 6 2 7...
Page 142: ...142 176 SIGRIST PHOTOMETER AG u u u RMA u u u u u RMA RMA...
Page 173: ...SIGRIST PHOTOMETER AG 173 176...
Page 174: ...174 176 SIGRIST PHOTOMETER AG...
Page 175: ...SIGRIST PHOTOMETER AG 175 176...