Mounting
SIGRIST-PHOTOMETER AG
21 / 176
Incorrect mounting may result in the formation
of deposits or accumulation of air bubbles. The
devices must not be mounted standing (up-
wards) or hanging (downwards).
The photometer can be mounted with a stan-
dardised in-line housing or a flange connection
with a special measuring cell, both in horizontal
and vertical process lines. In the vertical fitting
position, the cable penetration must point
downwards. For horizontal fitting, the cable
penetration is on the side.
3.2
Installation on VARINLINE
®
connector
u
Mount the photometer
(1)
including seal
(7)
with clamp ring
(6)
on the VARINLINE
®
con-
nector.
u
Ensure that the groove
(2)
points in the
flow direction.
u
Mount the blanking plate
(3)
, including
seal
(4)
with clamp ring
(5)
on the VARIN-
LINE
®
connector.
2
6
1
3
5
4
7
3.3
Fitting with flange connection
u
Fit the special measuring cell
(3)
in the
process line according to the dimension
drawing.
u
Fasten the photometer
(1)
crosswise to the
special measuring cell
(3)
with 4 screws
(4)
.
u
Tighten the screws
(4)
(tightening torque
min. 30 Nm, max. 35 Nm).
u
Ensure that the groove
(2)
points in the
flow direction.
1
3
4
2
3.4
Connecting the cooling unit
Use commercially available silicone hoses (interior diameter 6 mm) for the cooling unit.
DE
EN
FR
ES
NL
PT
ZH
RU
CS
JA
IT
PL
Summary of Contents for TurBiScat 2 Ex
Page 88: ...88 176 SIGRIST PHOTOMETER AG l l l l l l l l l l l l l 1 4 u 1 5 1 6...
Page 94: ...94 176 SIGRIST PHOTOMETER AG 4 3 0 2 mm2 100 m 10 5 TOUCH WiFi 5 1 l 0 5s l 1 2s 5 cm 6 u u...
Page 99: ...SIGRIST PHOTOMETER AG 99 176 u u u RMA u u u u u RMA RMA DE EN FR ES NL PT ZH RU CS JA IT PL...
Page 112: ...112 176 SIGRIST PHOTOMETER AG u u 5 u 4 3 u 3 u 1 u 6 u 7 u u 2 1 3 4 5 6 2 7...
Page 142: ...142 176 SIGRIST PHOTOMETER AG u u u RMA u u u u u RMA RMA...
Page 173: ...SIGRIST PHOTOMETER AG 173 176...
Page 174: ...174 176 SIGRIST PHOTOMETER AG...
Page 175: ...SIGRIST PHOTOMETER AG 175 176...