Servicing
26 / 176
SIGRIST-PHOTOMETER AG
The servicing interval reflects normal use.
The servicing interval is based on experience and is intended as a recommendation. It increases the fail-safety of
the system. The recommended maintenance cycle does not provide information about the durability of the compo-
nents.
Depending on the operating and ambient conditions, the interval must be shortened accordingly.
When
Who
What
Why
Annual/
"Humidity"
warning
Operator
Replace desiccant and seals
Maintaining measuring accuracy
and protecting the electronics
Annual
Operator
Clean sensor head
Maintaining measuring accuracy
Annual
Operator
Check calibration
Maintaining measuring accuracy
Annual
Operator
Replace VARINLINE® housing seals or
flange connection seals
Tightness at process line
Every 10 years
Service-
technician
Replace photometer battery
NOTICE!
It is mandatory to use lithium batter-
ies type CR1025 from the manufac-
turer "renata batteries".
Function servicing
8.2
Replace desiccant
NOTE
Condensation inside the electronics
If cold medium flows through the process line, humidity can condense when the process line is opened and dam-
age the electronics.
u
Only open the photometer when the medium temperature is
≥
room temperature.
If the desiccant needs to be replaced frequently, have the tightness checked by a service technician.
u
DANGER!
Explosion hazard!
Disconnect the service
voltage and disconnect all conductors.
u
Loosen Allen screw
(5)
.
u
Remove the protective sleeve
(4)
from the
clamp ring
(3)
.
u
Loosen and remove the clamp ring
(3)
.
u
Remove basic unit
(1)
from the sensor head.
u
Replace seal
(6)
.
u
Replace desiccant
(7)
.
u
Reassemble the device immediately in re-
verse order.
u
Note the alignment of the groove with
pin
(2)
.
1
3
4
5
6
2
7
Summary of Contents for TurBiScat 2 Ex
Page 88: ...88 176 SIGRIST PHOTOMETER AG l l l l l l l l l l l l l 1 4 u 1 5 1 6...
Page 94: ...94 176 SIGRIST PHOTOMETER AG 4 3 0 2 mm2 100 m 10 5 TOUCH WiFi 5 1 l 0 5s l 1 2s 5 cm 6 u u...
Page 99: ...SIGRIST PHOTOMETER AG 99 176 u u u RMA u u u u u RMA RMA DE EN FR ES NL PT ZH RU CS JA IT PL...
Page 112: ...112 176 SIGRIST PHOTOMETER AG u u 5 u 4 3 u 3 u 1 u 6 u 7 u u 2 1 3 4 5 6 2 7...
Page 142: ...142 176 SIGRIST PHOTOMETER AG u u u RMA u u u u u RMA RMA...
Page 173: ...SIGRIST PHOTOMETER AG 173 176...
Page 174: ...174 176 SIGRIST PHOTOMETER AG...
Page 175: ...SIGRIST PHOTOMETER AG 175 176...