Servicing
Instruction Manual OilGuard 2 Ex
13003E/3
81
WORKSTEP
ADDITIONAL INFO / IMAGES
6.
Position the new lamp holder so that it is
aligned in the direction of the arrow (see fig-
ure).
Ensure that the lamp holder is aligned
correctly on the pins.
7.
Fasten the lamp holder in place with the two
screws.
8.
Position the connector (A) over the socket (B),
push it upwards and fasten in place with a
quarter-turn clockwise.
Ensure that the pin is locked securely in
the groove on the connector (circle in figure
below).
9.
If soiled, remove the quartz glass from the
lamp cover and clean it.
Reinsert the glass back into the lamp cover af-
ter cleaning is completed.
10. Fasten the lamp cover in place with the two
screws.
11. Close the door according to Section 5.3.
12. Re-establish the service voltage.
13. Carry out recalibration according to
Section 9.3.