Servicing
Instruction Manual OilGuard 2 Ex
13003E/3
79
WORKSTEP
ADDITIONAL INFO / IMAGES
5.
Reinsert the protective glasses in the holder (B)
after cleaning is completed.
Ensure that the protective glasses (D)
are mounted as shown in pos. E.
A: Measuring cell wall
B: Holder for protective glass
C: Ring for protective glass
D: Protective glass
E: Mounted protective glass
6.
Check the inlet pipe for deposits and clean
when necessary.
Do not remove the deposits with a
knife or similar objects.
7.
Reinsert the inlet pipe and fasten in place with
the two locking screws.
8.
Reattach the cover on the measuring cell.
9.
Reattach the sample feed and return.