Troubleshooting
Instruction Manual AquaScat S
13798E/3
99
The following prioritized fault messages can be displayed:
PRIO MESSAGE
DESCRIPTION
POSSIBLE CAUSES
DEFAULT VALUES
The default values were load-
ed.
If no parameters were initial-
ized or if all parameters were
lost, the default values are
loaded.
CRC EXPERTS
A fault was determined when
the expert data was checked.
Electromagnetic malfunctions.
Defect in the electronic system.
CRC USER
A fault was determined when
the user data was checked.
Electromagnetic malfunctions.
Defect in the electronic system.
CRC DISPLAY
A fault was determined when
the display data was checked.
Electromagnetic malfunctions.
Defect in the electronic system.
EXT RAM
A fault was determined when
the RAM in the graphic con-
troller was checked.
Defect in the electronic system.
SW VERS
Software which is unsuitable
for this instrument type was
loaded.
Faulty software update.
Service technician
Table 5: Possible prioritized fault messages