Settings
Instruction Manual FireGuard 2
13542E/5
57
8.4
Setting the recalibration
If the
Auto start recal.
parameter is activated, then the installation of the checking unit in
the photometer automatically triggers a recalibration. This allows servicing duties to be car-
ried out without the use of a control unit. The state of the recalibration can be monitored via
the LED display (Section 7.2).
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Recalibration
button.
4.
Press the
General
button.
5.
Set
Auto start recal.
to
Active
or
Inactive
.
6.
Press the
Meas
button.
The instrument is in measuring
mode again.
8.5
Setting the Profibus DP parameters
This setting only has to be carried out if the optional Profibus module is used.
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Digi. interf.
button.
4.
Press the
Profibus DP
button.
5.
In the
Control
menu, select either
Local
or
External
.
▪
Local:
Values can only be read
via Profibus.
▪
External:
Values can be read
and written via Profibus.
6.
Enter the desired number under
Slave no.
.
Values between 1 and 240 are
possible.
7.
Press the
Meas
button.
The instrument is in measuring
mode again.
8.
To activate the parameters, the instrument
must be switched off and then switched on
again.